用于连续测量组织氧气摄取的高灵敏度设备。

High-sensitivity device for continuous measurements of tissue O2 uptake.

作者信息

Biron R, Dittmar A

出版信息

J Appl Physiol Respir Environ Exerc Physiol. 1977 Aug;43(2):370-4. doi: 10.1152/jappl.1977.43.2.370.

Abstract

An analysis system for continuous measurements of O2 uptake by rat diaphragm during several hours is described. The tissue was set in a chamber where a medium was flowing at a known and constant flow rate, temperature, and partial pressure of oxygen (PO2). The decrease in PO2 of the medium was measured with a Clark O2 cathode at the outlet of the chamber. To increase the sensitivity of the system and the signal-to-noise ratio of the measurements, a sweeping device was designed. It provided homogenization of the medium in the chamber. The main characteristics of this analysis system are a) a sensitivity of +/-0.2% of the mean PO2 value of the perfused medium, +/-0.8% of the mean O2 uptake of rat diaphragm, b) a linearity of +/-1.75% in the range of 0+/-200 Torr, c) a stability of +/-0.38% of the initial PO2 value, and d) a time constant of 8 min 30 s. This type of system can be used for high-sensitivity measurements of O2 uptake of any tissue after adaptation of the chamber and of the medium flow rate.

摘要

本文描述了一种用于连续数小时测量大鼠膈肌氧气摄取量的分析系统。将组织置于一个腔室中,在该腔室中,一种介质以已知且恒定的流速、温度和氧分压(PO2)流动。在腔室出口处,用克拉克氧电极测量介质中PO2的降低情况。为了提高系统的灵敏度和测量的信噪比,设计了一种扫描装置。它使腔室内的介质均匀化。该分析系统的主要特性为:a)对灌注介质平均PO2值的灵敏度为±0.2%,对大鼠膈肌平均氧气摄取量的灵敏度为±0.8%;b)在0±200托范围内的线性度为±1.75%;c)初始PO2值的稳定性为±0.38%;d)时间常数为8分30秒。在调整腔室和介质流速后,这种类型的系统可用于对任何组织的氧气摄取进行高灵敏度测量。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验