Lee M E, Chaux A, Matloff J M
J Thorac Cardiovasc Surg. 1977 Sep;74(3):433-5.
Endocardial pacemaker electrodes rely in part upon endocardial fixation proximal to the electrode tip to prevent ejection of the tip from the ventricular apex. Fixation of these electrodes to the superior vena cava and tricuspid valve, in particular, has been reported. Infection of endocardial electrodes necessitates their removal. This report concerns avulsion of a tricuspid valve leaflet during traction on an infected electrode. With the availability of new methods of apical fixation, the utilization of electrode sheathing materials which discourage endocardial fixation would increase the safety of their removal under the circumstances reported herein.
心内膜起搏器电极部分依赖于电极尖端近端的心内膜固定,以防止尖端从心室尖弹出。尤其是,已有报道称这些电极可固定于上腔静脉和三尖瓣。心内膜电极感染后必须将其移除。本报告涉及在牵拉感染电极时三尖瓣小叶撕脱的情况。随着新的心尖固定方法的出现,使用不利于心内膜固定的电极护套材料将增加在本文所述情况下移除电极的安全性。