Department of Nuclear Medicine, University of Massachusetts Medical Center, Worcester 01655, USA.
J Nucl Cardiol. 1995 Nov-Dec;2(6):513-24. doi: 10.1016/s1071-3581(05)80044-3.
Attenuation is believed to be one of the major causes of false-positive cardiac single-photon emission computed tomographic (SPECT) perfusion images. This article reviews the physics of attenuation, the artifacts produced by attenuation, and the need for scatter correction in combination with attenuation correction. The review continues with a comparison of the various configurations for transmission imaging that could be used to estimate patient specific attenuation maps, and an overview of how these are being developed for use on multiheaded SPECT systems, including discussions of truncation, noise, and spatial resolution of the estimated attenuation maps. Ways of estimating patient specific attenuation maps besides transmission imaging are also discussed.