Fell J W, Rayfield J M, Gulbin J P, Gaffney P T
School of Health Sciences, Griffith University - Gold Coast, Queensland, Australia.
Int J Sports Med. 1998 Apr;19(3):199-204. doi: 10.1055/s-2007-971904.
An investigation of the Accusport Lactate Analyser (ACC) was undertaken to assess its accuracy, linearity, reliability, and versatility. Accuracy was determined by comparing 32 blood assays on the ACC with those obtained using the Kodak Ektachem E250 (E250). Linearity was determined over the range 1.2-18.7 mmol/L. Reliability was measured at 1.7 (low) and 14.4 (high) mmol/L concentrations. Versatility involved determining the optimal sample volume, influence of haematocrit, and the effect of time delay from sampling to assay. Assays determined using the ACC confirm good agreement with the E250. The difference between the two methods compared for lactate determination does not exceed 1.1 mmol/L for concentrations up to approximately 20 mmol/L. Linearity is demonstrated by the regression equation y = 0.026 + 0.970x, r = 0.995. Reliability is expressed by the coefficient of variation of 7% (low) and 4.6% (high). The optimal sample size for accurate assay was determined to be between 20 microL and 50 microL. Mean lactate value did not vary significantly between a haematocrit of 35% and 45%. The Accusport is accurate, linear up to at least 18.7 mmol/L, has good reliability at high and low concentrations, and is able to analyse whole blood up to 15 minutes after sample application to the reagent strip.
对Accusport乳酸分析仪(ACC)进行了一项研究,以评估其准确性、线性、可靠性和通用性。通过将ACC上的32次血液检测结果与使用柯达Ektachem E250(E250)获得的结果进行比较来确定准确性。在1.2 - 18.7 mmol/L范围内确定线性。在1.7(低)和14.4(高)mmol/L浓度下测量可靠性。通用性包括确定最佳样本体积、血细胞比容的影响以及从采样到检测的时间延迟的影响。使用ACC进行的检测结果与E250高度一致。对于乳酸测定,两种方法之间的差异在浓度高达约20 mmol/L时不超过1.1 mmol/L。回归方程y = 0.026 + 0.970x,r = 0.995证明了线性。可靠性由7%(低)和4.6%(高)的变异系数表示。准确检测的最佳样本量确定为20微升至50微升之间。血细胞比容在35%至45%之间时,平均乳酸值无显著差异。Accusport准确,线性至少可达18.7 mmol/L,在高浓度和低浓度下均具有良好的可靠性,并且能够在将样本应用于试剂条后15分钟内分析全血。