Backofen U, Matysik F M, Hoffmann W, Ache H J
Forschungszentrum Karlsruhe, Institut für Instrumentelle Analytik, Germany.
Biomed Chromatogr. 1998 May-Jun;12(3):145-6. doi: 10.1002/(SICI)1099-0801(199805/06)12:3<145::AID-BMC788>3.0.CO;2-E.
Capillary batch injection analysis (CBIA) is applied to the determination of ascorbic acid in nanolitre samples at neutral pH. It is demonstrated that pulsed amperometric detection is superior to constant-potential amperometric detection with respect to signal stability. The calibration plot is linear within the studied concentration range between 0.5 mM and 5 mM and the reproducibility is characterized by a relative standard deviation of 0.80% for 10 successive injections of 188 nL of 3.5 mM ascorbic acid. The attainable sample throughput is up to 500 per hour.