Figeys D, Zhang Y, Aebersold R
Department of Molecular Biotechnology, University of Washington, Seattle, USA.
Electrophoresis. 1998 Oct;19(13):2338-47. doi: 10.1002/elps.1150191314.
We have previously described the use of a solid phase extraction (SPE) - capillary zone electrophoresis (CZE) - tandem mass spectrometry (MS/MS) system for protein analysis at the low femtomole to subfemtomole level. Here we describe the systematic optimization of a number of parameters which facilitate the use of the SPE-CZE-MS/MS system and further enhance its performance. Specifically, we describe a robust SPE cartridge design which can be assembled without the use of glue, the evaluation of procedures to chemically modify the inner wall of the fused-silica capillaries used in the system to improve separation and reproducibility, and the comparison of different reverse-phase (RP) resins used for the SPE cartridge. We also explored the effects of transient isotachophoresis with respect to system performance and compatibility with different fused-silica surface coatings, the RP resins used, and MS/MS. The enhanced performance of the optimized system is demonstrated by the analysis of calibrated tryptic digests of bovine serum albumin (BSA).
我们之前描述过使用固相萃取(SPE)-毛细管区带电泳(CZE)-串联质谱(MS/MS)系统在低飞摩尔至亚飞摩尔水平进行蛋白质分析。在此,我们描述了一系列参数的系统优化,这些参数有助于SPE-CZE-MS/MS系统的使用并进一步提高其性能。具体而言,我们描述了一种无需使用胶水即可组装的坚固型SPE柱管设计,评估了对系统中使用的熔融石英毛细管内壁进行化学修饰以改善分离和重现性的方法,以及对用于SPE柱管的不同反相(RP)树脂的比较。我们还探讨了瞬态等速电泳对系统性能的影响以及与不同熔融石英表面涂层、所用RP树脂和MS/MS的兼容性。通过对牛血清白蛋白(BSA)的校准胰蛋白酶消化产物进行分析,证明了优化系统的性能得到了增强。