Parker D L, Goodrich K C, Alexander A L, Buswell H R, Blatter D D, Tsuruda J S
Department of Radiology, University of Utah, Salt Lake City 84108-1218, USA.
Magn Reson Med. 1998 Dec;40(6):873-82. doi: 10.1002/mrm.1910400613.
In this study, the problem of small vessel visualization in magnetic resonance angiography is addressed. The loss of vessel contrast due to slow flow-related signal saturation can be compensated by the T1 reduction obtained from the use of an MR contrast agent, such as Gd-DTPA. The vessel/background signal-difference-to-noise ratio (SDNR) is shown to strongly depend on the imaging parameters, as well as on the time course of the blood T1 values obtained from the contrast injection. Specifically, it was found that vessel SDNR increases almost linearly with TR, if the sampling bandwidth is reduced proportionately.
在本研究中,探讨了磁共振血管造影中小血管可视化的问题。由于与血流相关的信号饱和导致的血管对比度丧失,可以通过使用磁共振造影剂(如钆喷酸葡胺)获得的T1缩短来补偿。血管/背景信号差噪比(SDNR)显示强烈依赖于成像参数,以及造影剂注射后血液T1值随时间的变化过程。具体而言,发现如果采样带宽按比例降低,血管SDNR几乎随重复时间(TR)呈线性增加。