We assessed the BioFocus 2000 capillary electrophoresis instrument for use in a routine clinical laboratory. We examined 210 serum samples received for serum protein electrophoresis by four methods: (1) The Bio-Rad HR015EC high-resolution serum protein kit on the BioFocus; (2) the Jenkins-Guerin (JG) method on the Applied Biosystems 270A HT Capillary Electrophoresis System (JG-ABI); (3) the Jenkins-Guerin method using the BioFocus (JG-BF); and (4) the quantitation of monoclonal bands found in 76 of the 210 samples was assayed by Helena Titan Hi-Res agarose gel electrophoresis (HRAGE). The correlation coefficient between the three sets of capillary electrophoresis monoclonal band results and the Helena quantitation was 0.92 or better. Although the quantitative comparison of monoclonal bands by HR015EC was very good, the lack of sharpness of monoclonal bands using the HR015EC kit meant our preference was to use the JG method on either the ABI or on the Biofocus.
我们评估了BioFocus 2000毛细管电泳仪在常规临床实验室中的使用情况。我们通过四种方法检测了210份用于血清蛋白电泳的血清样本:(1) BioFocus上的Bio-Rad HR015EC高分辨率血清蛋白试剂盒;(2) 应用生物系统公司270A HT毛细管电泳系统上的詹金斯-格林(JG)方法(JG-ABI);(3) 使用BioFocus的詹金斯-格林方法(JG-BF);(4) 对210份样本中的76份样本中发现的单克隆条带进行定量分析,采用海伦娜泰坦高分辨率琼脂糖凝胶电泳(HRAGE)。三组毛细管电泳单克隆条带结果与海伦娜定量分析之间的相关系数为0.92或更高。虽然使用HR015EC对单克隆条带进行定量比较非常好,但使用HR015EC试剂盒时单克隆条带缺乏清晰度,这意味着我们更倾向于在ABI或Biofocus上使用JG方法。