Frequency response of fluid-filled catheter-micromanometer systems used for measurement of left ventricular pressure.

作者信息

Scruggs V, Pietras R J, Rosen K M

出版信息

Am Heart J. 1975 May;89(5):619-24. doi: 10.1016/0002-8703(75)90508-6.

Abstract

The resonant frequencies of five commonly used fluid-filled catheters connected to a solidstate microdisplacement pressure gauge were 18 to 33 per cent higher than those obtained with a conventional manometer. Four catheters had a flat amplitude response to 26 Hz. or more at the 20 per cent amplitude distortion level. The dynamic response characteristic of certain fluid-filled catheters used with microdisplacement pressure gauges allows one to record high fidelity left ventricular pressure curves.

摘要

连接到固态微位移压力计的五种常用充液导管的共振频率比使用传统压力计测得的共振频率高18%至33%。在20%幅度失真水平下,有四种导管在26赫兹及以上频率具有平坦的幅度响应。与微位移压力计一起使用的某些充液导管的动态响应特性使人们能够记录高保真的左心室压力曲线。