Tuytten R, Lemière F, Dongen W Van, Esmans E L, Slegers H
Department of Chemistry, Nucleoside Research and Mass Spectrometry Unit, University of Antwerp, Groenenborgerlaan 171, B-2020 Antwerp, Belgium.
Rapid Commun Mass Spectrom. 2002;16(12):1205-15. doi: 10.1002/rcm.704.
A liquid chromatography/mass spectrometry (LC/MS) method for the analysis of complex mixtures of nucleoside mono-, di- and triphosphates has been developed. A short capillary column (35mm x 0.3mm i.d.) was operated under ion-pair high-performance liquid chromatography conditions and hyphenated to (negative) electrospray (tandem) mass spectrometry. As such, the separation of 12 nucleotides was performed by a binary gradient elution using CH(3)OH/H(2)O and N,N-dimethylhexylamine (N,N-DMHA) as ion-pairing agent. The influence of different N,N-DMHA concentrations on the chromatographic and mass spectrometric performance was evaluated to achieve optimal LC/MS conditions. In addition it was demonstrated that a controlled admission of ammonium dihydrogen phosphate (NH(4)H(2)PO(4)) improved both chromatographic performance and mass spectrometric detection. Because the system was hyphenated to an orthogonal designed electrospray interface (Z-spraytrade mark), long acquisition times were possible without loss of sensitivity.