Criterion validity study of the cervical range of motion (CROM) device for rotational range of motion on healthy adults.
作者信息
Tousignant Michel, Smeesters Cécil, Breton Anne-Marie, Breton Emilie, Corriveau Hélène
机构信息
Research Center on Aging, Sherbrooke Geriatric University Institute, Faculty of Medicine, Québec, Canada.
出版信息
J Orthop Sports Phys Ther. 2006 Apr;36(4):242-8. doi: 10.2519/jospt.2006.36.4.242.
STUDY DESIGN
This study compared range of motion (ROM) measurements using a cervical range of motion device (CROM) and an optoelectronic system (OPTOTRAK).
OBJECTIVES
To examine the criterion validity of the CROM for the measurement of cervical ROM on healthy adults.
BACKGROUND
Whereas measurements of cervical ROM are recognized as part of the assessment of patients with neck pain, few devices are available in clinical settings. Two papers published previously showed excellent criterion validity for measurements of cervical flexion/extension and lateral flexion using the CROM.
METHODS AND MEASURES
Subjects performed neck rotation, flexion/extension, and lateral flexion while sitting on a wooden chair. The ROM values were measured by the CROM as well as the OPTOTRAK.
RESULTS
The cervical rotational ROM values using the CROM demonstrated a good to excellent linear relationship with those using the OPTOTRAK: right rotation, r = 0.89 (95% confidence interval, 0.81-0.94), and left rotation, r = 0.94 (95% confidence interval, 0.90-0.97). Similar results were also obtained for flexion/extension and lateral flexion ROM values.
CONCLUSION
The CROM showed excellent criterion validity for measurements of cervical rotation. We propose using ROM values measured by the CROM as outcome measures for patients with neck pain.
研究设计
本研究比较了使用颈椎活动度测量仪(CROM)和光电系统(OPTOTRAK)测量的活动范围(ROM)。
目的
检验CROM在健康成年人中测量颈椎ROM的标准效度。
背景
尽管颈椎ROM测量被认为是颈部疼痛患者评估的一部分,但临床环境中可用的设备很少。此前发表的两篇论文显示,使用CROM测量颈椎屈伸和侧屈具有出色的标准效度。
方法和测量
受试者坐在木椅上进行颈部旋转、屈伸和侧屈。ROM值由CROM以及OPTOTRAK测量。
结果
使用CROM测量的颈椎旋转ROM值与使用OPTOTRAK测量的值显示出良好至出色的线性关系:右旋,r = 0.89(95%置信区间,0.81 - 0.94),左旋,r = 0.94(95%置信区间,0.90 - 0.97)。屈伸和侧屈ROM值也获得了类似结果。
结论
CROM在测量颈椎旋转方面显示出出色的标准效度。我们建议将CROM测量的ROM值用作颈部疼痛患者的结局指标。