J Appl Clin Med Phys. 2021 Jun;22(6):262-267. doi: 10.1002/acm2.13282. Epub 2021 May 28.
This study aims to investigate the influence of the phantom position on weighted computed tomography dose index (CTDI ) in cone beam computed tomography (CBCT) when assuming breast irradiation. Computed tomography dose index (CTDI) was measured by the x-ray volume imaging of CBCT using parameters for image-guided radiation therapy (IGRT) in right breast irradiation. The measurement points of CTDI ranged from 0 (center) to 16 cm in the right-left (RL) direction, and from 0 (center) to 7.5 cm in the anterior-posterior (AP) direction, which assumed right breast irradiation. A nonuniform change exists in the relative value of CTDI when the phantom deviated from the isocenter of CBCT. The CTDI was ~30% lower compared with the value at the isocenter of CBCT when the phantom deviated 7.5 and 16 cm at the AP and RL directions, respectively. This study confirmed the influence of the phantom position on the CTDI values of CBCT. The CTDI measured at the isocenter of CBCT overestimates that measured at the irradiation center of the breast.
本研究旨在探讨在假设乳腺癌放射治疗时,体模位置对锥形束 CT(CBCT)加权 CT 剂量指数(CTDI)的影响。采用图像引导放射治疗(IGRT)的 CBCT 射线容积成像来测量 CT 剂量指数(CTDI)。在右侧乳腺照射时,CTDI 的测量点范围从 0(中心)到 16cm 的左右(RL)方向,以及从 0(中心)到 7.5cm 的前后(AP)方向。当体模偏离 CBCT 的等中心时,CTDI 的相对值存在不均匀变化。当体模在 AP 和 RL 方向上分别偏离 7.5cm 和 16cm 时,CTDI 比 CBCT 等中心处的值低约 30%。本研究证实了体模位置对 CBCT 的 CTDI 值的影响。在 CBCT 的等中心处测量的 CTDI 高估了在乳腺照射中心处测量的 CTDI。