Wada M, Kinoshita S, Itayama Y, Kuroda N, Nakashima K
School of Pharmaceutical Sciences, Nagasaki University, Japan.
J Chromatogr B Biomed Sci Appl. 1999 Jan 22;721(2):179-86. doi: 10.1016/s0378-4347(98)00499-x.
A sensitive HPLC method for the determination of phenol and chlorophenols was developed. The fluorescence labeling reaction of phenols with 4-(4,5-diphenyl-1H-imidazol-2-yl)benzoyl chloride (DIB-Cl) was completed in 30 min at 60 degrees C. The separation of DIB-derivatives of five representative phenols, i.e., phenol, o-, p-chlorophenol, 2,4-dichlorophenol, 2,4,6-trichlorophenol, was achieved within 35 min with an ODS column using isocratic elution. The detection limits of these DIB derivatives at a signal-to-noise ratio (SIN) of 3 were in the range of 0.024 to 0.08 microM (0.12-0.45 pmol/20 microl injection). Twelve kinds of DIB derivatives with phenols containing mono-, di-, tri-, tetra- and penta-chlorophenol were also well separated within 208 min by changing the elution conditions. The derivatives were stable for at least for 24 h when they were placed at room temperature in the dark. The proposed method was applied to the assay of human urine samples and free and total phenol were determined. The relative standard deviations (RSDs) of the proposed method for within and between-day assay were <7.0% and <14.2%, respectively. The average concentrations of free and total phenol found in urine (n=6) were 4.3+/-2.5 and 29.5+/-14.0 microM, respectively.
建立了一种灵敏的高效液相色谱法用于测定苯酚和氯酚。酚类与4-(4,5-二苯基-1H-咪唑-2-基)苯甲酰氯(DIB-Cl)的荧光标记反应在60℃下30分钟内完成。使用ODS柱通过等度洗脱在35分钟内实现了五种代表性酚类(即苯酚、邻氯苯酚、对氯苯酚、2,4-二氯苯酚、2,4,6-三氯苯酚)的DIB衍生物的分离。这些DIB衍生物在信噪比(S/N)为3时的检测限在0.024至0.08微摩尔范围内(0.12 - 0.45皮摩尔/20微升进样量)。通过改变洗脱条件,十二种含单、二、三、四和五氯苯酚的酚类的DIB衍生物在208分钟内也得到了很好的分离。当这些衍生物在室温黑暗条件下放置时,至少在24小时内是稳定的。所提出的方法应用于人体尿液样本的分析,测定了游离酚和总酚。所提出方法的日内和日间分析的相对标准偏差(RSD)分别<7.0%和<14.2%。尿液中(n = 6)游离酚和总酚的平均浓度分别为4.3±2.5和29.5±14.0微摩尔。