Cutaneous pO2-measurements in newborns with respiratory distress syndrome.

作者信息

Bossi E, Breitenstein M, Lenzin B

出版信息

Helv Paediatr Acta. 1976 Dec;31(4-5):335-45.

PMID:1017982
Abstract

Correlation between cutaneously measured pO2-values and arterial pO2 in 9 newborns with respiratory distress syndromes was investigated with the Basel skin-electrode [3]. At an electrode temperature of 42 degrees C, correlation was low (r = 0.63), at 43 degrees C the coefficient increased to 0.71 at 44 degrees C the over all statistical correlation improved markedly (r = 0.97). cpO2 for given paO2-values, however, varied widely: reliable discrimination between hypo-, normo- and hyperoxemia was impossible. Continuous registration allows recognition of the consequences of common clinical situations and nursing manipulations on cpO2. The electrode, however, is still susceptible of technical modifications.

摘要

采用巴塞尔皮肤电极[3]研究了9例患有呼吸窘迫综合征的新生儿经皮测量的pO2值与动脉pO2之间的相关性。在电极温度为42℃时,相关性较低(r = 0.63),在43℃时系数增至0.71,在44℃时总体统计相关性显著改善(r = 0.97)。然而,对于给定的paO2值,cpO2变化很大:无法可靠地区分低氧血症、正常氧血症和高氧血症。连续记录有助于识别常见临床情况及护理操作对cpO2的影响。然而,该电极仍易于进行技术改进。