Suppr超能文献

一项使用单根导管记录肌电图、阻抗和压力变化来研究输尿管蠕动的研究。

A study of ureteric peristalsis using a single catheter to record EMG, impedance, and pressure changes.

作者信息

Roshani H, Dabhoiwala N F, Tee S, Dijkhuis T, Kurth K H, Ongerboer de Visser B W, de Jong J M, Lamers W H

机构信息

Department of Urology, Academic Medical Center, Faculty of Medicine, Amsterdam, The Netherlands.

出版信息

Tech Urol. 1999 Mar;5(1):61-6.

Abstract

Ureteric peristalsis transports a urinary bolus from the renal pelvis to the bladder. We developed an intraluminal catheter with a pressure transducer on it to study intraluminal pressure changes and a twin bipolar electrode to record the ureteric EMG and impedance (Z) changes during a peristaltic wave. Five female New Yorkshire pigs (50-60 kg) were studied under light halothane anesthesia (5% at induction/1% for maintenance). A steady state of hydration was maintained using intravenous saline infusion. EMG spike burst activity was studied at a 10-cm interval using low (0-30) Hz filters. Impedance between the same electrodes is measured simultaneously in higher frequencies (1-5 KHz) as a function of ureteric motor activity. Pressure generation in the ureteric lumen was also measured simultaneously by a transducer on the same catheter. A digital signal processing program (Poly 4.9) was used for analysis. Parenteral furosemide was used to induce diuresis. Resting ureteric impedance (Z(R)) decreases to Z(B) (Z bolus) during the passage of the urinary bolus. Passage of a contractile zone during a peristaltic wave increases impedance from Z(B) to its Z(R) level and initiates a pressure rise. Bolus length (the length Z(B)) is not constant and decreases distally. EMG corresponds well in time to impedance. Z(R) disappears after infusion of furosemide because of increased urine load and changes of intraluminal ionic environment. The contractile segment of a ureteric peristaltic wave appears to be represented by an elevated Z segment (Z(C)). Pressure rise is recorded only at the beginning of a contractile zone. A specially adapted intraluminal catheter can be used to study peristalsis in the upper urinary tract. One can study all the three components of ureteric peristalsis (excitation, contraction, and intraluminal pressure rise) using such a catheter.

摘要

输尿管蠕动将尿团从肾盂输送至膀胱。我们研发了一种管腔内带有压力传感器的导管,用于研究管腔内压力变化,还配备了双极电极来记录输尿管肌电图(EMG)以及蠕动波期间的阻抗(Z)变化。在轻度氟烷麻醉(诱导时5%/维持时1%)下对5只雌性新约克夏猪(50 - 60千克)进行研究。通过静脉输注生理盐水维持水合稳态。使用低(0 - 30)Hz滤波器,每隔10厘米研究一次EMG尖峰爆发活动。同时在较高频率(1 - 5千赫兹)下测量相同电极之间的阻抗,作为输尿管运动活动的函数。同一导管上的传感器同时测量输尿管腔内的压力产生。使用数字信号处理程序(Poly 4.9)进行分析。静脉注射速尿以诱导利尿。尿团通过期间,静息输尿管阻抗(Z(R))降至Z(B)(Z团块)。蠕动波期间收缩区的通过使阻抗从Z(B)增加到其Z(R)水平并引发压力升高。团块长度(Z(B)的长度)不恒定且向远端减小。EMG在时间上与阻抗很好地对应。输注速尿后,由于尿负荷增加和管腔内离子环境变化,Z(R)消失。输尿管蠕动波的收缩段似乎由升高的Z段(Z(C))表示。仅在收缩区开始时记录到压力升高。一种经过特殊改装的管腔内导管可用于研究上尿路的蠕动。使用这样的导管可以研究输尿管蠕动的所有三个组成部分(兴奋、收缩和管腔内压力升高)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验