使用多环表面线圈的磁共振波谱分析。

MR spectroscopy using multi-ring surface coils.

作者信息

King S B, Ryner L N, Tomanek B, Sharp J C, Smith I C

机构信息

Institute for Biodiagnostics, National Research Council of Canada, Winnipeg, Canada.

出版信息

Magn Reson Med. 1999 Oct;42(4):655-64. doi: 10.1002/(sici)1522-2594(199910)42:4<655::aid-mrm6>3.0.co;2-d.

Abstract

A spatially uniform B(1)-field is preferred for MR imaging and spectroscopy. Unfortunately, volume coils are sometimes unavailable, or do not provide adequate RF power or SNR for some applications. In quantitative MRS, mean metabolite concentration cannot be evaluated when the coil response is nonuniform, unless an assumption is made concerning the metabolite spatial distribution. It is well known that standard single-loop surface coils, although offering high SNR characteristics, have poor B(1) homogeneity. New multi-ring surface coils are proposed which produce a locally uniform B(1) field, with sensitivity and power requirements comparable to those of standard surface coils. MR spectroscopy using two and three-ring versions of this "local volume coil" result in spatial localization essentially identical to that obtained with a volume coil but with much improved RF power and SNR characteristics. When compared to standard surface coils, the multi-ring coil offers much improved water suppression and localization, as well as reduced outer voxel contamination, with only a small loss in SNR and moderate increase in SAR. In summary, the multi-ring coil operates midway between the volume coil and the standard surface coil, retaining the most advantageous properties of both. Magn Reson Med 42:655-664, 1999. Published 1999 Wiley-Liss, Inc.

摘要

在磁共振成像和光谱学中,空间均匀的B(1)场是优选的。不幸的是,体线圈有时不可用,或者对于某些应用不能提供足够的射频功率或信噪比。在定量磁共振波谱分析中,当线圈响应不均匀时,除非对代谢物的空间分布做出假设,否则无法评估平均代谢物浓度。众所周知,标准的单环表面线圈虽然具有高信噪比特性,但B(1)均匀性较差。本文提出了一种新型的多环表面线圈,它能产生局部均匀的B(1)场,其灵敏度和功率要求与标准表面线圈相当。使用这种“局部体线圈”的两环和三环版本进行磁共振波谱分析,其空间定位与体线圈基本相同,但射频功率和信噪比特性有了很大改善。与标准表面线圈相比,多环线圈在水抑制和定位方面有了很大改进,外部体素污染也有所减少,只是信噪比略有损失,比吸收率适度增加。总之,多环线圈在体线圈和标准表面线圈之间起到了中间作用,保留了两者最有利的特性。《磁共振医学》42:655 - 664,1999年。1999年由威利 - 利斯公司出版。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验