de Bruijn A J, Tange R A, Dreschler W A
Department of Otolaryngology-Head and Neck Surgery, Academic Medical Center, University of Amsterdam, The Netherlands.
Am J Otol. 1999 Sep;20(5):573-80.
Evaluation of hearing results after implantation of a Teflon piston (type Causse; Xomed Surgical Products, Jacksonville, FL) and of a pure gold piston (K-piston; Heinz Kurz GmbH Medizintechnik, Dusslingen, Germany), both with a shaft diameter of 0.4 mm in cases of otosclerosis requiring stapedotomy. An important difference between both prostheses is the difference in mass: the gold piston is three times heavier than the Teflon piston.
Retrospective analyses were carried out of the presurgery and postsurgery audiologic results obtained after primary stapedotomy by implantation of 62 Teflon pistons and 66 gold pistons. The results were compared according to mean values of several audiometric parameters. Furthermore, individual audiometric results have been evaluated with the "Amsterdam Hearing Evaluation Plots" (AHEPs) as a method for visual presentation of hearing results from each operated ear. With this method, "unsuccessfully" and "successfully" inserted prostheses can be recognized easily and a more realistic comparison between prostheses is possible.
It was found that in the overall group the heavier gold prosthesis gives a significantly larger gain in air-conductive hearing at 2 kHz (p < 0.05) and in the speech frequency range 0.5 to 2 kHz (p < 0.05). There were no significant intergroup differences with regard to change in bone-conduction and air-bone gaps. Analysis of the hearing results of the subgroup that included only the "successfully" implanted prostheses according to the criteria of the AHEPs had mainly consequences for improvement of air-conduction thresholds: none of the intergroup differences were statistically significant.
For a fair comparison between prostheses, it is necessary to take only the prostheses into account that are inserted properly and that are functioning under normal conditions with regard to transmission of sound vibrations. After analysis of the hearing results of these "successfully" implanted prostheses, a trend was noticed that the heavier gold piston gives more gain in the low- and mid-frequency range and the light-weighted Teflon piston gives more gain in the high-frequency range, although none of the differences were significant.
评估在镫骨切除术治疗耳硬化症的病例中,植入特氟龙活塞(Causse型;Xomed Surgical Products公司,佛罗里达州杰克逊维尔)和纯金活塞(K型活塞;Heinz Kurz GmbH Medizintechnik公司,德国杜斯林根)后的听力结果。两种假体的一个重要区别在于质量差异:金活塞比特氟龙活塞重三倍。
对62个特氟龙活塞和66个金活塞植入初次镫骨切除术后的术前和术后听力学结果进行回顾性分析。根据几个听力测定参数的平均值比较结果。此外,个体听力测定结果已通过“阿姆斯特丹听力评估图”(AHEPs)进行评估,作为一种直观呈现每只手术耳听力结果的方法。通过这种方法,可以轻松识别“植入失败”和“植入成功”的假体,并且可以对假体进行更实际的比较。
发现在整个组中,较重的金假体在2 kHz的气导听力(p < 0.05)以及在0.5至2 kHz的语音频率范围内(p < 0.05)有显著更大的增益。在骨导和气骨导间距的变化方面,组间没有显著差异。根据AHEPs标准对仅包括“植入成功”假体的亚组的听力结果进行分析,主要影响气导阈值的改善:组间差异均无统计学意义。
为了对假体进行公平比较,有必要仅考虑正确植入且在声音振动传递方面正常运作的假体。在分析这些“植入成功”假体的听力结果后,注意到一种趋势,即较重的金活塞在低频和中频范围内增益更多,而轻质的特氟龙活塞在高频范围内增益更多,尽管差异均不显著。