Defibrillator analyzers automate the inspection and preventive maintenance (IPM) testing of defibrillators. They need to be able to test at least four basic defibrillator performance characteristics: discharge energy, synchronized-mode operation, automated external defibrillation, and ECG monitoring. We prefer that they also be able to test a defibrillator's external noninvasive pacing function--but this is not essential if a facility already has a pacemaker analyzer that can perform this testing. In this Evaluation, we tested seven defibrillator analyzers from six suppliers. All seven units accurately measure the energies of a variety of discharge wave-forms over a wide range of energy levels--from 1 J for use in a neonatal intensive care unit to 360 J for use on adult patients requiring maximum discharge energy. Most of the analyzers are easy to use. However, only three of the evaluated units could perform the full range of defibrillator tests that we prefer. We rated these units Acceptable--Preferred. Three more units could perform four of the five tests, they could not test the pacing feature of a defibrillator. These units were rated Acceptable. The seventh unit could perform only discharge energy testing and synchronized-mode testing and was difficult to use. We rate that unit Acceptable--Not Recommended.