Cavagna R, Tessarin C, Tarroni G, Casol D, De Silvestro L, Fabbian F
Nephrology and Dialysis Unit, Belluno Hospital, Italy.
Perit Dial Int. 1999 Nov-Dec;19(6):540-3.
Peritoneal catheter displacement appears to be related to various causes including omental attachment, bowel contractions, peritoneal adhesions, and the catheter floating in the dialysis fluid. In order to prevent this complication, which can impair peritoneal dialysis efficacy, Di Paolo et al. designed a "self-locating catheter" (SLC) that is similar to the Tenckhoff catheter (TC) and includes a small tungsten cylinder (weight 12 g) at the distal end (Di Paolo N, et al. The self positioning catheter. Proceedings of the VII Italian Congress on Peritoneal Dialysis. Milan: Wichtig Editore, 1993:539-42). The weight of the tip prevents the catheter from floating and migrating by gravitation toward the Douglas cavity.
Starting in 1996, we implanted SLC in 15 continuous ambulatory peritoneal dialysis patients. The aim of our study was to compare the ratio of migration and other catheter-related complications between patients with SLC (196 patient-months) and 13 patients with conventional straight TC (295 patient-months).
Peritonitis ratio, catheter complication rate, dialysate inflow and outflow, and weekly creatinine clearance were similar in the two groups of patients. The incidence of catheter displacement was significantly higher (p = 0.0349) in theTC group than in the SLC group (4 vs 0).
In our experience, the SLC seems to be useful in preventing catheter migration by continuous gravitation of its extremity toward the pelvic cavity.
腹膜导管移位似乎与多种原因有关,包括网膜附着、肠道收缩、腹膜粘连以及导管在透析液中漂浮。为了预防这种可能损害腹膜透析疗效的并发症,迪·保罗等人设计了一种“自定位导管”(SLC),它类似于Tenckhoff导管(TC),在远端包含一个小钨圆柱体(重量12克)(迪·保罗N等人。自定位导管。第七届意大利腹膜透析大会论文集。米兰:维希蒂出版社,1993:539 - 42)。尖端的重量可防止导管因重力作用漂浮并向Douglas腔迁移。
从1996年开始,我们在15例持续性非卧床腹膜透析患者中植入了SLC。我们研究的目的是比较SLC患者(196患者 - 月)和13例传统直型TC患者(295患者 - 月)之间的导管移位率及其他与导管相关的并发症。
两组患者的腹膜炎发生率、导管并发症发生率、透析液流入和流出情况以及每周肌酐清除率相似。TC组的导管移位发生率显著高于SLC组(4比0,p = 0.0349)。
根据我们的经验,SLC似乎有助于通过其末端持续向盆腔重力作用来防止导管迁移。