Supko J G, Nair R V, Seiden M V, Lu H
Division of Hematology-Oncology, Massachusetts General Hospital and Harvard Medical School, Boston 02114-2689, USA.
J Pharm Biomed Anal. 1999 Dec;21(5):1025-36. doi: 10.1016/s0731-7085(99)00229-0.
An improved method for assaying paclitaxel in human plasma by high-performance liquid chromatography (HPLC) with UV detection at 227 nm has been developed by adapting previously reported sample preparation methods and chromatographic conditions to facilitate semi-automated sample cleanup using a column switching technique. Manual sample manipulations were limited to isolating the drug and internal standard from plasma (1.0 ml) by liquid-liquid extraction using tert-butyl methyl ether. The sample extract was initially loaded onto a short cartridge column containing a cyanopropyl stationary phase. During the predetermined time interval that the drug and internal standard eluted from the cartridge, 1.50-2.20 min, a fully automated 6-position switching valve was used to direct the effluent onto an octylsilica analytical column. The same mobile phase, composed of acetonitrile-methanol-ammonium acetate buffer (pH 5.0; 20 mM) (76:19:105, v/v/v) and delivered at flow rate of 1.0 ml/min, was used for both separations. The overall retention times of paclitaxel and the internal standard were 10.9 and 18.1 min, respectively. The analytical method was thoroughly validated for quantitating paclitaxel in plasma at concentrations ranging from 6 to 586 nM (5-500 ng/ml). The lowest concentration of paclitaxel measured with acceptable day-to-day accuracy (100.2%) and precision (RSD 11.7%, n = 21, 5 months) was 6 nM (5 ng/ml). The sensitivity and selectivity of the assay proved to be more than adequate for monitoring steady-state plasma concentrations of the drug when administered to cancer patients as a 96 h continuous intravenous infusion in combination with other anticancer agents, such as doxorubicin and topotecan. Moreover, the heart-cutting procedure prevented the problematic introduction of interfering nonpolar plasma components onto the analytical column, thereby enhancing sample throughput while decreasing the technical demands of the assay. The method was found to be extremely reproducible and robust during extended use for the routine analysis of plasma specimens acquired from several clinical trials.
通过采用先前报道的样品制备方法和色谱条件,开发了一种改进的高效液相色谱(HPLC)法,用于在227nm处进行紫外检测,以测定人血浆中的紫杉醇,该方法采用柱切换技术促进半自动样品净化。手动样品操作仅限于通过使用叔丁基甲醚的液液萃取从血浆(1.0ml)中分离药物和内标。样品提取物首先加载到含有氰丙基固定相的短柱上。在药物和内标从柱中洗脱的预定时间间隔(1.50 - 2.20分钟)内,使用全自动6位切换阀将流出物引导至辛基硅胶分析柱上。两种分离均使用相同的流动相,由乙腈 - 甲醇 - 醋酸铵缓冲液(pH 5.0;20mM)(76:19:105,v/v/v)组成,流速为1.0ml/min。紫杉醇和内标的总保留时间分别为10.9分钟和18.1分钟。该分析方法在定量血浆中浓度范围为6至586 nM(5 - 500 ng/ml)的紫杉醇时经过了全面验证。以可接受的日常准确度(100.2%)和精密度(RSD 11.7%,n = 21,5个月)测量的紫杉醇最低浓度为6 nM(5 ng/ml)。当与其他抗癌药物(如阿霉素和拓扑替康)联合作为96小时连续静脉输注给药给癌症患者时,该测定法的灵敏度和选择性被证明足以监测药物的稳态血浆浓度。此外,中心切割程序可防止有问题的非极性血浆干扰成分进入分析柱,从而提高样品通量,同时降低测定法的技术要求。在对从多个临床试验获得的血浆标本进行常规分析的长期使用过程中,发现该方法具有极高的重现性和稳健性。