Li X A, Wang R, Yu C, Suntharalingam M
Department of Radiation Oncology, University of Maryland, Baltimore, MD 21201-1595, USA.
Int J Radiat Oncol Biol Phys. 2000 Mar 1;46(4):1043-9. doi: 10.1016/s0360-3016(99)00457-5.
Both beta and gamma emitters are currently used in the catheter-based intravascular brachytherapy. The dosimetric effects due to the presence of metallic stents and calcified plaques have not been fully addressed. This work compares these effects for two most commonly used beta and gamma sources ( (90)Sr and (192)Ir).
An EGS4 Monte Carlo package was used to calculate dose in water for a (90)Sr (supplied by NOVOSTE) and an (192)Ir (Supplied by BEST) source, with or without the presence of a calcified plaque or a metallic stent. Plaques of different shape (shell and disk), size and density, and two types of stainless-steel stents (ring or mesh stent) were studied. The ring stent consists of identical rings stacked along the long axis of the sources. The gap between two rings is 0.3 mm. The mesh stents are made of identical square (0.1 x 0.1 or 0.2 x 0.2 mm(2)) holes separated from each other by stainless-steel wire. The cross section of wire for both ring and mesh stents is 0.1 x 0.1 mm(2). A dose perturbation factor (DPF), defined as the ratio of the doses with and without the presence of a plaque or a stent, was introduced to quantify the effects. A carefully chosen set of EGS4 transport parameters for the small geometry in question was used in the calculation.
The radial and axial dose distributions calculated in water were found to agree with the published measurements to within 3%. The dose perturbations due to the presence of calcified plaques or metallic stents were found far more significant for the (90)Sr source than those for the (192)Ir source. Up to 30% dose reduction behind a plaque were observed for the (90)Sr source, while the dose reduction for the (192)Ir source was found to be negligible. The dose enhancement inside a plaque was as high as 10% for the beta source or 6% for the gamma source. In the presence of a stent, the DPF was in the range of 1.15-0.75 for the beta source, while it was almost equal to 1.0 for the gamma source.
The dose perturbation due to the presence of a calcified plaque or a metallic stent is significant for the beta source. The dose reduction in the region beyond a plaque or a stent could be more than 20%. For the gamma source, the dose effect behind a plaque or a stent is practically negligible. These dosimetric differences between the beta and gamma sources in the presence of a calcified plaque or metallic stent should be considered in the dose prescription of intravascular brachytherapy.
目前,β射线和γ射线发射源均应用于基于导管的血管内近距离放射治疗。金属支架和钙化斑块的存在所产生的剂量学影响尚未得到充分研究。本研究比较了两种最常用的β射线和γ射线源(锶-90和铱-192)的这些影响。
使用EGS4蒙特卡罗程序包计算在有或没有钙化斑块或金属支架存在的情况下,锶-90(由NOVOSTE提供)源和铱-192(由BEST提供)源在水中的剂量。研究了不同形状(壳形和盘形)、大小和密度的斑块,以及两种类型的不锈钢支架(环形或网状支架)。环形支架由沿源的长轴堆叠的相同环组成。两个环之间的间隙为0.3毫米。网状支架由相同的正方形(0.1×0.1或0.2×0.2平方毫米)孔组成,这些孔由不锈钢丝彼此隔开。环形和网状支架的钢丝横截面均为0.1×0.1平方毫米。引入剂量扰动因子(DPF),定义为有斑块或支架存在与不存在时的剂量之比,以量化这些影响。在计算中使用了为所讨论的小几何形状精心选择的一组EGS4传输参数。
发现在水中计算的径向和轴向剂量分布与已发表的测量结果相符,误差在3%以内。发现钙化斑块或金属支架的存在对锶-90源产生的剂量扰动比对铱-192源的剂量扰动要显著得多。对于锶-90源,观察到斑块后方剂量降低高达30%,而铱-192源的剂量降低可忽略不计。β射线源在斑块内的剂量增强高达10%,γ射线源为6%。在有支架的情况下,β射线源的DPF在1.15 - 0.75范围内,而γ射线源的DPF几乎等于1.0。
钙化斑块或金属支架的存在对β射线源产生的剂量扰动显著。斑块或支架后方区域的剂量降低可能超过20%。对于γ射线源,斑块或支架后方的剂量效应实际上可忽略不计。在血管内近距离放射治疗的剂量处方中应考虑钙化斑块或金属支架存在时β射线源和γ射线源之间的这些剂量学差异。