Kawashima I, Ohno H, Sachdeva R
Department of Dental Materials Science, School of Dentistry, Health Sciences University of Hokkaido, Japan.
Dent Mater J. 1999 Dec;18(4):403-12. doi: 10.4012/dmj.18.403.
Simple three point bending tests were performed on Ni-Ti wires with three different Af points (1 degree C, 13 degrees C and 34 degrees C) to clarify the relationship between Af temperature and load changes under constant deformation. Each wire was deformed at 37 degrees C and then thermal changes were imposed by temperatures of 2 degrees C or 60 degrees C. The load changes with thermal changes from 37 degrees C to 2 degrees C or 60 degrees C showed the same tendency on the wires with different Af points: In the loading stage, the load became lower than the initial level at 37 degrees C and in the unloading stage, the load became higher than the initial load. The largest load change in the unloading stage was measured with the 13 degrees C Af point wire. Care must be taken when handling Ni-Ti wire with an Af point of less than 1 degree C in order to prevent it from reaching the limit of critical stress of slip deformation when the temperature in the mouth rises to above 40 degrees C.
对具有三个不同奥氏体终了(Af)点(1℃、13℃和34℃)的镍钛丝进行了简单的三点弯曲试验,以阐明在恒定变形下Af温度与载荷变化之间的关系。每根丝在37℃下变形,然后通过2℃或60℃的温度施加热变化。从37℃到2℃或60℃的热变化引起的载荷变化在具有不同Af点的丝上呈现相同趋势:在加载阶段,载荷低于37℃时的初始水平,而在卸载阶段,载荷高于初始载荷。卸载阶段最大的载荷变化是在Af点为13℃的丝上测得的。处理Af点低于1℃的镍钛丝时必须小心,以防止当口腔温度升至40℃以上时,丝达到滑移变形临界应力的极限。