Jackson P E, Gokhale S, Streib T, Rohrer J S, Pohl C A
Dionex Corporation, Sunnyvale, CA 94088-3606, USA.
J Chromatogr A. 2000 Aug 4;888(1-2):151-8. doi: 10.1016/s0021-9673(00)00557-4.
Ammonium perchlorate, a key ingredient in solid rocket propellants, has been found in ground and surface waters in a number of U.S. states, and perchlorate contamination of public drinking water wells is now a serious problem in California. Perchlorate poses a health risk and preliminary data from the U.S. EPA reports that exposure to less than 4-18 microg/l provides adequate human health protection. An improved ion chromatographic method was developed for the determination of low microg/l levels of perchlorate in ground and drinking waters based on a Dionex IonPac AS16 column, an hydroxide eluent generated using an EG40 automated eluent generator, large loop (1000 microl) injection, and suppressed conductivity detection. The method is free of interferences from common inorganic anions, linear over the range of 2-100 microg/l perchlorate, and quantitative recoveries are obtained for low microg/l levels of perchlorate in spiked ground and drinking water samples. The MDL of 150 ng/l permits quantification of perchlorate below the levels that ensure adequate health protection.
高氯酸铵是固体火箭推进剂的关键成分,在美国多个州的地表水和地下水中都被检测到,目前加利福尼亚州公共饮用水井的高氯酸盐污染已成为一个严重问题。高氯酸盐对健康构成风险,美国环境保护局的初步数据表明,接触低于4 - 18微克/升的高氯酸盐可确保人体健康。基于戴安公司的IonPac AS16色谱柱、使用EG40自动淋洗液发生器产生的氢氧化物淋洗液、大定量环(1000微升)进样以及抑制型电导检测,开发了一种改进的离子色谱法,用于测定地表水和饮用水中低微克/升水平的高氯酸盐。该方法不受常见无机阴离子的干扰,在2 - 100微克/升高氯酸盐范围内呈线性,加标的地表水和饮用水样品中低微克/升水平的高氯酸盐定量回收率良好。150纳克/升的方法检出限能够对低于确保健康所需水平的高氯酸盐进行定量。