Dowling R D, Etoch S W, Stevens K, Butterfield A, Koenig S E, Johnson A, Chiang B, Gray L A
Department of Surgery, University of Louisville and the Jewish Hospital Heart and Lung Institute, Kentucky, USA.
ASAIO J. 2000 Sep-Oct;46(5):579-81. doi: 10.1097/00002480-200009000-00014.
Potential benefits of heart transplantation are limited by the severe donor organ shortage. The AbioCor implantable replacement heart has been developed as a potential alternative to heart transplantation. We report our initial experience with the AbioCor in a bovine model. A right thoracotomy was performed for access to the heart and great vessels. After initiation of cardiopulmonary bypass, excision of the native ventricles was followed by orthotopic placement of the IRH and complete implantation of the transcutaneous energy transfer coil, controller, and battery pack. Invasive monitoring of IVC, SVC, carotid artery, pulmonary artery, and left atrial (LA) pressures was performed in all animals. Twelve calves have undergone implantation of the AbioCor. There were three early deaths, one from bleeding, one from respiratory failure, and one from neurodysfunction from low flow during CPB. Nine animals have had a normal recovery and survived a mean of 24.5 days (range, 4-48 days). All the animals have demonstrated excellent hemodynamics with the maintenance of normal pressures in the LA, SVC, IVC, pulmonary artery, and aorta. Adjustment of the right-sided internal hydraulic fluid shunt has allowed for control of right-left balance and, thereby, manipulation of left and right side filling pressures. Late morbidity has consisted of neck wound infection and sepsis, pneumonia, and bleeding. Successful orthotopic implantation of all components of the AbioCor has been achieved in a bovine model. This device has demonstrated restoration of normal hemodynamics and excellent function of the atrial hydraulic shunt to achieve right-left balance.
心脏移植的潜在益处因供体器官严重短缺而受限。已研发出阿比奥科(AbioCor)可植入式人工心脏作为心脏移植的一种潜在替代方案。我们报告在牛模型中使用阿比奥科的初步经验。通过右侧开胸术暴露心脏和大血管。启动体外循环后,切除天然心室,然后将可植入式人工心脏原位植入,并完全植入经皮能量传输线圈、控制器和电池组。对所有动物进行了下腔静脉、上腔静脉、颈动脉、肺动脉和左心房压力的有创监测。12头小牛接受了阿比奥科植入。有3例早期死亡,1例死于出血,1例死于呼吸衰竭,1例死于体外循环期间低流量导致的神经功能障碍。9只动物恢复正常,平均存活24.5天(范围4 - 48天)。所有动物均表现出良好的血流动力学,左心房、上腔静脉、下腔静脉、肺动脉和主动脉压力维持正常。通过调整右侧内部液压分流装置可控制左右平衡,从而调节左右心室充盈压力。晚期并发症包括颈部伤口感染和败血症、肺炎及出血。在牛模型中已成功实现阿比奥科所有组件的原位植入。该装置已证明可恢复正常血流动力学,且心房液压分流功能良好,可实现左右平衡。