Lo I K, King G J, Patterson S D, Johnson J A, Chess D G
Bioengineering Research Laboratory, London, Ontario, Canada.
J Hand Surg Br. 2001 Feb;26(1):22-4. doi: 10.1054/jhsb.2000.0481.
This study evaluated the compressive capabilities of the 3.0 mm Synthes cannulated screw and threaded washer. A transverse osteotomy was performed at the waist of eight cadaveric scaphoids and a custom-designed load cell was inserted before internal fixion with a 3.0 mm cannulated screw and threaded washer. The mean intrascaphoid compression achieved was 108 (SD, 60) N. This compressive force is comparable to that produced by standard cortical screws.