Suppr超能文献

对应用于MM50跑道型电子回旋加速器的动态多叶准直器所产生的调强射束稳定性进行测试。

Testing of the stability of intensity modulated beams generated with dynamic multileaf collimation, applied to the MM50 racetrack microtron.

作者信息

Dirkx M L, Heijmen B J

机构信息

University Hospital Rotterdam/Daniel den Hoed Cancer Center, Department of Radiation Oncology, Rotterdam, The Netherlands.

出版信息

Med Phys. 2000 Dec;27(12):2701-7. doi: 10.1118/1.1326450.

Abstract

Recently, we have published a method for the calculation of required leaf trajectories to generate optimized intensity modulated x-ray beams by means of dynamic multileaf collimation [Phys. Med. Biol. 43, 1171-1184 (1998)]. For the MM50 Racetrack Microtron it has been demonstrated that the dosimetric accuracy of this method, in combination with the dose calculation algorithm of the Cadplan 3D treatment planning system, is adequate for a clinical application (within 2% or 0.2 cm). Prior to initiating patient treatment with dynamic multileaf collimation (DMLC), tests have been performed to investigate the stability of DMLC fields generated at the MM50, (i) in time, (ii) subject to gantry rotation and (iii) in case of treatment interrupts, e.g., caused by an error detected by the treatment machine. The stability of relative dose profiles, normalized to a reference point in a relatively flat part of the modulated beam profile, was assessed from measurements with an electronic portal imaging device (EPID), with a linear diode array attached to the collimator and with film. The dose in the reference point was monitored using an ionization chamber. Tests were performed for several intensity modulated fields using 10 and 25 MV photon beams. Based on film measurements for sweeping 0.1 cm leaf gaps it was concluded that in an 80 days period the variation in leaf positioning was within 0.05 cm, without requiring any recalibration. For a uniform 10x10 cm2 field, realized dynamically by a scanning 0.4x10 cm2 slit beam, a maximum variation in slit width of 0.01 cm was derived from ionization chamber measurements, both in time and for gantry rotation. For a clinical example, the dose in the reference point reproduced within 0.2% (1 SD) over a period of 100 days. Apart from regions with very large dose gradients, variations in the relative beam profiles measured with the EPID were generally less than 1% (1 SD). For different gantry angles the dose profiles also reproduced within 1%, showing that gravity has a negligible influence. No significant deviations between uninterrupted and interrupted treatments could be observed, indicating that the effects of acceleration and deceleration of the leaves are negligible and that a DMLC treatment can be finished correctly after a treatment interrupt. Our previous and present studies have demonstrated that the dosimetric accuracy and stability of intensity modulated beams, generated at the MM50 by means of dynamic multileaf collimation, are adequate for clinical use. Patient treatment using dynamic multileaf collimation has been started in our clinic.

摘要

最近,我们发表了一种通过动态多叶准直来计算所需叶轨迹以生成优化调强X射线束的方法[《物理医学与生物学》43, 1171 - 1184 (1998)]。对于MM50跑道型电子回旋加速器,已经证明该方法的剂量学精度与Cadplan 3D治疗计划系统的剂量计算算法相结合,足以用于临床应用(在2%或0.2厘米范围内)。在开始使用动态多叶准直(DMLC)进行患者治疗之前,已进行测试以研究在MM50产生的DMLC射野的稳定性,(i)随时间的稳定性,(ii)在机架旋转情况下的稳定性,以及(iii)在治疗中断的情况下,例如由治疗机器检测到的错误导致的中断。通过电子射野成像装置(EPID)、连接到准直器的线性二极管阵列以及胶片测量,对归一化到调强射野轮廓相对平坦部分中的参考点的相对剂量分布的稳定性进行了评估。使用电离室监测参考点处的剂量。使用10和25 MV光子束对多个调强射野进行了测试。基于对0.1厘米叶间隙扫描的胶片测量得出,在80天的时间段内,叶定位的变化在0.05厘米以内,无需任何重新校准。对于通过扫描0.4×10平方厘米的狭缝束动态实现的均匀10×10平方厘米射野,从电离室测量得出,狭缝宽度在时间和机架旋转方面的最大变化为0.01厘米。对于一个临床实例,参考点处的剂量在100天的时间段内再现精度在0.2%(1标准差)以内。除了剂量梯度非常大的区域外,用EPID测量的相对射野轮廓的变化通常小于1%(1标准差)。对于不同的机架角度,剂量分布也在1%以内再现,表明重力的影响可忽略不计。未观察到连续治疗和中断治疗之间有显著偏差,这表明叶片加速和减速的影响可忽略不计,并且在治疗中断后DMLC治疗可以正确完成。我们之前和目前的研究表明,通过动态多叶准直在MM50产生的调强射野的剂量学精度和稳定性足以用于临床。我们诊所已开始使用动态多叶准直进行患者治疗。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验