Malik N, Gunn J, Shepherd L, Crossman D C, Cumberland D C, Holt C M
Cardiovascular Research Group, Department of Medicine, Clinical Sciences Center, Northern General Hospital, Herries Road, Sheffield S5 7AU, United Kingdom.
J Invasive Cardiol. 2001 Mar;13(3):193-201.
To examine the angiographic (quantitative coronary angiography), morphometric, light microscopic (LM) (i.e., histology and immunohistochemical staining) and electron microscopic (EM) findings after implantation of phosphorylcholine (PC)-coated compared to uncoated stents in porcine coronary arteries.
Forty (25 PC-coated, 15 uncoated) divYsio stents were implanted into the coronary arteries of 20 pigs. Quantitative coronary angiography (QCA) was performed pre-stent and post-implantation in fifteen pigs, at 28 days. Two pigs were killed at 5 days (LM and scanning EM), one pig at 14 days (scanning EM) and 17 pigs at 28 days (LM, scanning EM, transmission EM). At 28 days, thirty-two of 34 stented segments excised were formalin-fixed, of which 30 were embedded in resin and sectioned for morphometry and LM. Remaining stents were examined by TEM and SEM.
No angiographically occlusive thrombosis occurred in any of the stents. LM at 5 days showed endothelialization of PC-coated and uncoated stents, which was also confirmed by scanning EM at 14 days. At 28 days, QCA and morphometry showed no significant differences between PC-coated and uncoated stents. A few inflammatory cells were seen in both stent types at 5 days but there was no inflammatory or additional tissue reaction to PC-coated compared to uncoated stents at 28 days.
The divYsio stents, with or without PC coating, performed equally well in terms of acute patency, 28-day QCA and morphometry. The PC coating allows a stent to endothelialize normally and is not associated with specific histological changes. The PC coating on the divYsio stent appears biocompatible.
研究在猪冠状动脉中植入磷酰胆碱(PC)涂层支架与未涂层支架后的血管造影(定量冠状动脉造影)、形态学、光学显微镜(LM)(即组织学和免疫组织化学染色)及电子显微镜(EM)检查结果。
将40个(25个PC涂层,15个未涂层)divYsio支架植入20头猪的冠状动脉。15头猪在植入前和植入后28天进行定量冠状动脉造影(QCA)。2头猪在5天时处死(用于LM和扫描EM),1头猪在14天时处死(用于扫描EM),17头猪在28天时处死(用于LM、扫描EM、透射EM)。在28天时,切除的34个带支架节段中的32个用福尔马林固定,其中30个包埋在树脂中用于形态学和LM检查。其余支架用TEM和SEM检查。
所有支架均未发生血管造影闭塞性血栓形成。5天时的LM显示PC涂层和未涂层支架均有内皮化,14天时的扫描EM也证实了这一点。28天时,QCA和形态学显示PC涂层和未涂层支架之间无显著差异。5天时两种支架类型均可见少量炎症细胞,但28天时与未涂层支架相比,PC涂层支架无炎症或额外的组织反应。
无论有无PC涂层,divYsio支架在急性通畅性、28天QCA和形态学方面表现同样良好。PC涂层使支架能够正常内皮化,且与特定的组织学变化无关。divYsio支架上的PC涂层似乎具有生物相容性。