Shou W Z, Jiang X, Beato B D, Naidong W
Bioanalytical Chemistry, Covance Laboratories Inc., 3301 Kinsman Boulevard, Madison, Wisconsin 53704, USA.
Rapid Commun Mass Spectrom. 2001;15(7):466-76. doi: 10.1002/rcm.255.
A high-throughput bioanalytical method based on automated sample transfer, automated solid phase extraction, and fast liquid chromatography/tandem mass spectrometry (LC/MS/MS) analysis, has been developed for the determination of the analgesic fentanyl in human plasma. Samples were transferred into 96-well plates using an automated sample handling system. Automated solid phase extraction (SPE) was carried out using a 96-channel programmable liquid-handling workstation using a mixed-mode sorbent. The extracted samples were then dried down, reconstituted and injected onto a silica column using an aqueous/organic mobile phase with tandem mass spectrometric detection. The method has been validated over the concentration range 0.05-100 ng/mL fentanyl in human plasma, based on a 0.25-mL sample size. The assay is sensitive, specific and robust. More than 2000 samples have been analyzed using this method. The automation of the sample preparation steps not only increased the analysis throughput, but also facilitated the transfer of the method between different bioanalytical laboratories of the same organization.
已开发出一种基于自动样品转移、自动固相萃取和快速液相色谱/串联质谱(LC/MS/MS)分析的高通量生物分析方法,用于测定人血浆中的镇痛剂芬太尼。使用自动样品处理系统将样品转移到96孔板中。使用96通道可编程液体处理工作站和混合模式吸附剂进行自动固相萃取(SPE)。然后将萃取的样品吹干、复溶,并使用含串联质谱检测的水/有机流动相注入硅胶柱。基于0.25 mL的样品量,该方法已在人血浆中芬太尼浓度范围为0.05 - 100 ng/mL内进行了验证。该测定法灵敏、特异且稳健。已使用此方法分析了2000多个样品。样品制备步骤的自动化不仅提高了分析通量,还便于该方法在同一机构的不同生物分析实验室之间转移。