Leyland M, Zinicola E, Bloom P, Lee N
Hillingdon Hospital NHS Trust, Uxbridge, UK.
Eye (Lond). 2001 Apr;15(Pt 2):202-5. doi: 10.1038/eye.2001.61.
Standard intraocular lenses (IOL) only correct the spherical component of aphakic refractive error. This study describes clinical experience with a foldable, injectable, toric IOL.
Keratometric, refractive and visual data were collected on patients listed for cataract surgery. Toric IOLs were offered where keratometric cylinder was greater than 1.5 dioptres. Small-incision phacoemulsification surgery was performed and the IOL implanted with its long axis along the steep corneal axis. Post-operative refractive data were compared with pre-operative and expected refraction using vector analysis software.
Results of 22 eyes of 16 consecutive patients implanted with toric IOLs are reported. Two IOLs rotated more than 30 degrees in the first 24 h and were re-dialled surgically. Two further IOLs (9%) rotated more than 30 degrees during follow-up. In 21 of 22 eyes (95%) the refractive astigmatism was reduced, with a mean 73% of planned correction (vector analysis) achieved.
The toric IOL is a useful surgical tool to reduce the refractive effects of pre-existing corneal astigmatism. Design modification to prevent IOL rotation would make results more predictable.
标准人工晶状体(IOL)仅矫正无晶状体屈光不正的球面成分。本研究描述了一种可折叠、可注射的环曲面IOL的临床经验。
收集白内障手术患者的角膜曲率、屈光和视力数据。当角膜曲率柱镜大于1.5屈光度时,提供环曲面IOL。行小切口超声乳化手术,将IOL长轴沿角膜陡峭轴植入。使用矢量分析软件将术后屈光数据与术前及预期屈光进行比较。
报告了连续16例患者22只眼植入环曲面IOL的结果。2只IOL在最初24小时内旋转超过30度,并通过手术重新调整。另外2只IOL(9%)在随访期间旋转超过30度。22只眼中的21只(95%)屈光性散光降低,平均达到计划矫正量的73%(矢量分析)。
环曲面IOL是一种有用的手术工具,可减少术前角膜散光的屈光效应。对IOL旋转进行设计改进将使结果更可预测。