In vitro study of a multi-layer piezoelectric crystal attic hearing implant.

作者信息

Mills R P, Wang Z G, Abel E W

机构信息

Department of Otolaryngology, Royal Infirmary, Edinburgh, UK.

出版信息

J Laryngol Otol. 2001 May;115(5):359-62. doi: 10.1258/0022215011907730.

Abstract

We have developed a prototype middle-ear hearing implant which uses a multilayer piezoelectric actuator. In this series of experiments the actuator was attached to the medial wall of the attic so that it makes contact with the body of the incus. Initial in vitro evaluation has been carried out using a laser vibrometer (Polytec CLV) to measure stapes velocity. Stapes displacement is calculated by mathematical integration. The device used in this way is particularly effective at transmitting high frequency sound to the stapes. When switched off the actuator impairs the transmission of sound to the ossicular chain at low frequencies, but this effect is only 7 dB at most. The stapes displacements resulting from the action of the implant have a linear relationship with the voltages used to drive the system. The high capacitance of the present actuator means that its power requirements are higher than that of other comparable devices. An optimal method of coupling the device to the incus has yet to be identified.

摘要

我们已经开发出一种使用多层压电致动器的中耳听力植入原型。在这一系列实验中,致动器附着在鼓室上壁的内侧壁,使其与砧骨体接触。最初的体外评估是使用激光测振仪(Polytec CLV)来测量镫骨速度。镫骨位移通过数学积分计算得出。以这种方式使用的该装置在将高频声音传输到镫骨方面特别有效。当关闭时,致动器会在低频时损害声音向听骨链的传输,但这种影响最多只有7分贝。由植入物作用产生的镫骨位移与用于驱动系统的电压呈线性关系。当前致动器的高电容意味着其功率需求高于其他类似装置。将该装置与砧骨耦合的最佳方法尚未确定。