Galuppo L D, Stover S M, Jensen D G, Willits N H
J. D. Wheat Veterinary Orthopedic Research Laboratory, School of Veterinary Medicine,University of California, Davis, CA, USA.
Vet Surg. 2001 Jul-Aug;30(4):332-40. doi: 10.1053/jvet.2001.24386.
To compare drilling, tapping, and screw-insertion torque, force, and time for the 4.5-mm AO and 6.5-mm Acutrak Plus (AP) bone screws, and to compare the mechanical shear strength and stiffness of a simulated complete lateral condylar fracture of the equine third metacarpal bone (MC3) stabilized with either an AO or AP screw.
In vitro biomechanical assessment of screw-insertion variables, and shear failure tests of a bone-screw-stabilized simulated lateral condylar fracture.
Eight pairs of cadaveric equine MC3s
Metacarpi were placed in a fixture and centered on a biaxial load cell in a materials-testing system to measure torque, compressive force, and time for drilling, tapping, and screw insertion. Standardized simulated lateral condylar fractures were stabilized by either an AO or AP screw and tested in shear until failure. A paired t test was used to assess differences between screws, with significance set at P < .05.
Insertion and mechanical shear testing variables were comparable for AO and AP insertion equipment and screws.
The 6.5-mm tapered AP screw can be inserted in equine third metacarpal condyles and is mechanically comparable with the 4.5-mm AO screw for fixation of a simulated lateral condylar fracture.
Considering the comparable mechanical behavior, the potential for less-persistent soft-tissue irritation with the headless design, and the ability to achieve interfragmentary compression by inserting the screw in one hole drilled perpendicular to the fracture plane, the 6.5-mm tapered AP screw may be an attractive alternative for repair of incomplete lateral condylar fractures in horses.
比较4.5毫米AO骨螺钉和6.5毫米Acutrak Plus(AP)骨螺钉的钻孔、攻丝及拧入扭矩、力量和时间,并比较用AO或AP螺钉固定马第三掌骨(MC3)模拟完全性外侧髁突骨折后的机械剪切强度和刚度。
对螺钉拧入变量进行体外生物力学评估,以及对骨螺钉固定的模拟外侧髁突骨折进行剪切破坏试验。
八对马尸体的MC3
将掌骨置于夹具中,并在材料测试系统中的双轴测力传感器上居中,以测量钻孔、攻丝和拧入螺钉的扭矩、压缩力和时间。用AO或AP螺钉固定标准化模拟外侧髁突骨折,并进行剪切试验直至破坏。采用配对t检验评估螺钉之间的差异,显著性设定为P < 0.05。
AO和AP的拧入设备及螺钉在拧入和机械剪切测试变量方面具有可比性。
6.5毫米锥形AP螺钉可拧入马第三掌骨髁突,在固定模拟外侧髁突骨折方面,其力学性能与4.5毫米AO螺钉相当。
考虑到力学性能相当、无头设计对软组织刺激持续时间可能较短,以及通过在垂直于骨折平面钻的一个孔中拧入螺钉实现骨折块间加压的能力,6.5毫米锥形AP螺钉可能是修复马不完全外侧髁突骨折的一个有吸引力的选择。