Ogrodnik P J, Moorcroft C I, Thomas P B
School of Engineering and Advanced Technology, Staffordshire University, Stafford, UK.
Proc Inst Mech Eng H. 2001;215(4):405-14. doi: 10.1243/0954411011535984.
This paper presents a new design for a device to monitor the motion of fracture fragments in diaphyseal tibial fractures. The device measures the motion that occurs at the fracture site when loaded by gait or by manipulation. It has undergone rigorous calibration and acceptance trials. The device has been used in ethically approved research clinics held at the North Staffordshire Hospital (40 patients). The paper presents a selection of results obtained using the new device. The results demonstrate several new ways of assessing fracture healing by examining fracture site motion. The following conclusions were drawn: 1. If fracture monitoring devices are to be attached to bone screws, it is essential to minimize bone screw errors. To do this, each patient must have similar bone screw lengths, orientations, alignment and siting. This is only achievable using a peroperative reduction device. 2. If fracture stiffness is to be used as a measure of fracture healing, load rate should be controlled; at the very least strain rate should be controlled. 3. It is imperative that fracture stiffness be measured in more than one plane by a biplanar device so that asymmetry may be accommodated. 4. Fracture stiffness, on its own, is probably not a sufficiently rigorous measure of healing end-point. The quantifiably viscoelastic properties of healing callus should be taken into account.
本文介绍了一种用于监测胫骨干骨折骨折碎片运动的装置的新设计。该装置可测量在步态加载或手法操作时骨折部位发生的运动。它已经过严格的校准和验收试验。该装置已在北斯塔福德郡医院举行的符合伦理规范的研究诊所中使用(40名患者)。本文展示了使用该新装置获得的一系列结果。这些结果证明了通过检查骨折部位运动来评估骨折愈合的几种新方法。得出了以下结论:1. 如果要将骨折监测装置附着于骨螺钉,必须尽量减少骨螺钉误差。为此,每位患者的骨螺钉长度、方向、对齐方式和位置必须相似。这只有使用术中复位装置才能实现。2. 如果要用骨折刚度作为骨折愈合的指标,应控制加载速率;至少应控制应变率。3. 必须使用双平面装置在多个平面测量骨折刚度,以便能够考虑到不对称性。4. 仅骨折刚度本身可能不是衡量愈合终点的足够严格的指标。应考虑愈合骨痂的可量化粘弹性特性。