Aaslid R, Levang O, Fröysaker T, Skagseth E, Hall K V
Scand J Thorac Cardiovasc Surg. 1975;9(2):81-4. doi: 10.3109/14017437509139179.
An index of the "in situ" forward flow characteristics of an aortic valve prosthesis is suggested. The effective area index, Ia, is defined as the ratio between the effective orifice area, Aeff, and the area occupied by the valve in the aortic root. Aeff is calculated from a hydraulic formula using peroperative measurements of instantaneous ascending aortic flow, Fao, and the pressure gradient, deltaP. Ia was found to be 0.29 +/- 0.33 (S.D.) for Lillehei-Kaster valves and 0.32 +/- 0.059 (S.D.) for Björk-Shiley valves. The variations in valve performance among the Björk-Shiley valves may partly be due to differences in valve orientation in the aortic root. Attention should therefore be given to the choice of the opening direction of the disc in order to optimize valve performance in the patient.
本文提出了一种用于评估主动脉瓣假体“原位”正向血流特性的指标。有效面积指数(Ia)定义为有效开口面积(Aeff)与主动脉根部瓣膜所占面积之比。Aeff通过水力公式,利用术中测量的升主动脉瞬时血流(Fao)和压力梯度(deltaP)来计算。结果发现,Lillehei-Kaster瓣膜的Ia为0.29±0.33(标准差),Björk-Shiley瓣膜的Ia为0.32±0.059(标准差)。Björk-Shiley瓣膜之间的性能差异可能部分归因于瓣膜在主动脉根部的方向不同。因此,为了使患者的瓣膜性能达到最佳,应注意选择瓣膜盘的开口方向。