Suppr超能文献

Compression depth estimation for CPR quality assessment using DSP on accelerometer signals.

作者信息

Aase Sven O, Myklebust Helge

机构信息

Stavanger University College, Department of Electrical and Computer Engineering, Norway.

出版信息

IEEE Trans Biomed Eng. 2002 Mar;49(3):263-8. doi: 10.1109/10.983461.

Abstract

Chest compression is a vital part of cardiopulmonary resuscitation (CPR). This paper demonstrates how the compression depth can be estimated using the principles of inertia navigation. The proposed method uses accelerometer sensors, one placed on the patient's chest, the other beside the patient. The acceleration-to-position conversion is performed using discrete-time digital signal processing (DSP). Instability problems due to integration are combated using a set of boundary conditions. The proposed algorithm is tested on a mannequin in harsh environments, where the patient is exposed to external forces as in a boat or car, as well as improper sensor/patient alignment. The overall performance is an estimation depth error of 4.3 mm in these environments, which is reduced to 1.6 mm in a regular, flat-floor controlled environment.