Short communication: anomalous image quality phantom scores in magnification mammography: evidence of phase contrast enhancement.

作者信息

Kotre C J, Birch I P, Robson K J

机构信息

Regional Medical Physics Department, Newcastle General Hospital, Newcastle-upon-Tyne NE4 6BE, UK.

出版信息

Br J Radiol. 2002 Feb;75(890):170-3. doi: 10.1259/bjr.75.890.750170.

Abstract

Anomalously high image quality scores were noted for images of the Leeds TORMAM phantom obtained using magnification mammography. Comparison of optical density profiles of fibre features in the images with non-magnified images and images previously obtained using an in-line phase contrast geometry showed the presence of phase contrast enhancement in the magnification images. The effect on the phantom score is particularly marked for this design of phantom owing to its use of fibres, which tend to enhance well. A large proportion of the phantom score is associated with fibrous features. It is concluded that direct comparison of TORMAM phantom scores from magnified images with those from non-magnified images is not valid due to the different balance of physical mechanisms forming the two kinds of image.

摘要

使用放大乳腺摄影术获得的利兹TORMAM体模图像的图像质量分数异常高。将这些图像中纤维特征的光密度分布与未放大图像以及先前使用在线相衬成像几何结构获得的图像进行比较,结果表明放大图像中存在相衬增强。由于该体模设计使用了易于增强效果的纤维,因此对体模分数的影响尤为显著。体模分数的很大一部分与纤维特征相关。得出的结论是,由于形成两种图像的物理机制平衡不同,因此直接比较放大图像和未放大图像的TORMAM体模分数是无效的。