Continuous ambulatory right heart pressure measurements with an implantable hemodynamic monitor: a multicenter, 12-month follow-up study of patients with chronic heart failure.

作者信息

Magalski Anthony, Adamson Philip, Gadler Frederick, Böehm Michael, Steinhaus David, Reynolds Dwight, Vlach Kathryn, Linde Cecilia, Cremers Bodo, Sparks Brandon, Bennett Tom

机构信息

Mid-America Heart Institute, St Luke's Hospital, Kansas City, MO 64111, USA.

出版信息

J Card Fail. 2002 Apr;8(2):63-70. doi: 10.1054/jcaf.2002.32373.

Abstract

BACKGROUND

We describe the performance of an implantable hemodynamic monitor (IHM) that allows continuous recording of heart rate, patient activity levels, and right ventricular systolic, right ventricular diastolic, and estimated pulmonary artery diastolic pressures. Pressure parameters derived from the implantable monitor were correlated to measurements made with a balloon-tipped catheter to establish accuracy and reproducibility over time in patients with chronic heart failure (CHF).

METHODS AND RESULTS

IHM devices were implanted in 32 patients with CHF (left ventricular ejection fraction, 29% +/- 11%; range, 14%-62%) and were tested with right heart catheterization at implantation and 3, 6, and 12 months later. Hemodynamic variables were digitally recorded simultaneously from the IHM and catheter. Values were recorded during supine rest, peak response of Valsalva maneuver, sitting, peak of a 2-stage (25-50 W) bicycle exercise test, and final rest period. The median of 21 paired beat-to-beat cardiac cycles was analyzed for each intervention. A total of 217 paired data values from all maneuvers were analyzed for 32 patients at implantation and 129 paired data values for 20 patients at 1 year. The IHM and catheter values were not different at baseline or at 1 year (P >.05). Combining all interventions, correlation coefficients were 0.96 and 0.94 for right ventricular systolic pressure, 0.96 and 0.83 for right ventricular diastolic pressure, and 0.87 and 0.87 for estimated pulmonary artery diastolic pressure at implantation and 1 year, respectively.

CONCLUSIONS

The IHM and a standard reference pressure system recorded comparable right heart pressure values in patients with CHF. This implantable pressure transducer is accurate over time and provides a means to precisely monitor the hemodynamic condition of patients with CHF in a continuous fashion.

摘要

背景

我们描述了一种植入式血流动力学监测器(IHM)的性能,该监测器能够连续记录心率、患者活动水平以及右心室收缩压、右心室舒张压和估计的肺动脉舒张压。将来自植入式监测器的压力参数与用球囊导管进行的测量结果相关联,以确定慢性心力衰竭(CHF)患者随时间推移的准确性和可重复性。

方法与结果

将IHM装置植入32例CHF患者(左心室射血分数为29%±11%;范围为14%-62%)体内,并在植入时以及3、6和12个月后进行右心导管检查测试。血流动力学变量由IHM和导管同时进行数字记录。在仰卧休息、瓦尔萨尔瓦动作的峰值反应、坐姿、两阶段(25-50瓦)自行车运动测试的峰值以及最后的休息期记录数值。对每次干预分析21个配对的逐搏心动周期的中位数。对32例患者植入时的所有动作共分析了217个配对数据值,对20例患者1年时分析了129个配对数据值。IHM和导管的值在基线时或1年时无差异(P>.05)。综合所有干预措施,植入时和1年时右心室收缩压的相关系数分别为0.96和0.94,右心室舒张压的相关系数分别为0.96和0.83,估计的肺动脉舒张压的相关系数分别为0.87和0.87。

结论

在CHF患者中,IHM和标准参考压力系统记录的右心压力值相当。这种植入式压力传感器随时间推移是准确的,并提供了一种以连续方式精确监测CHF患者血流动力学状况的方法。