Nelson Robert M, Stebor Andrea D, Groh Catherine M, Timoney Paula M, Theobald Karen S, Friedman Bruce A
Division of Neonatology, Department of Pediatrics, University of South Florida, Tampa 33606-3309, USA.
Blood Press Monit. 2002 Apr;7(2):123-9. doi: 10.1097/00126097-200204000-00006.
New software (SuperSTAT(R) algorithm) with enhancements aimed at shorter determination times was developed for a non-invasive blood pressure (NIBP) device and a clinical evaluation was conducted to verify accuracy.
To determine the accuracy of the new algorithm according to ANSI/AAMI SP10-1992 and SP10A-1996 American National Standard for Electronic or Automated Sphygmomanometers.
The blood pressure values obtained from the test device were compared to the intra-arterial blood pressure reference standard (IBP).
The NIBP and IBP comparisons for systolic, diastolic, and mean arterial pressure met the 1992 ANSI/AAMI accuracy standards by being within a mean difference of +/- 5 mmHg and standard deviation of < or = 8 mmHg.
Non-invasive blood pressure determinations taken with the new algorithm, developed to provide greater patient comfort due to faster determinations, were accurate when compared to neonatal IBP.