Nishinaka Tomohiro, Tatsumi Eisuke, Taenaka Yoshiyuki, Katagiri Nobumasa, Ohnishi Hiroyuki, Shioya Kyoko, Fukuda Toshihide, Oshikawa Mitsuo, Sato Koichi, Tsukiya Tomonori, Homma Akihiko, Takewa Yoshiaki, Takano Hisateru, Sato Masaki, Kashiwabara Susumu, Tanaka Hidenori, Sakai Kazunari, Matsuda Toshiaki
Department of Artificial Organs, National Cardiovascular Center, Osaka, Japan.
Artif Organs. 2002 Jun;26(6):548-51. doi: 10.1046/j.1525-1594.2002.06886_3.x.
We developed an extracorporeal membrane oxygenation (ECMO) system with high antithrombogenicity and durability characteristics for prolonged continuous cardiopulmonary support. The oxygenator consists of a special hollow-fiber-type polyolefin gas-exchange membrane, which has an ultrathin dense layer in contact with the blood, in order to prevent plasma leakage during protracted use (Platinum Cube NCVC). The centrifugal pump (RotaFlow) is free of seals. The entire blood-contacting surface of the system is coated with a newly developed heparin material (Toyobo-NCVC coating). We performed a venoarterial bypass in a goat, and the ECMO system was driven for 34 days without systemic anticoagulants. Plasma leakage from the oxygenator did not occur, and sufficient gas exchange performance was maintained. Thrombus formation was hardly observed in the ECMO system except in the casing margins of the oxygenator. This ECMO system showed potential for long-term cardiopulmonary support with minimal or no use of systemic anticoagulants.
我们开发了一种具有高抗血栓形成性和耐用性的体外膜肺氧合(ECMO)系统,用于长时间持续心肺支持。氧合器由一种特殊的中空纤维型聚烯烃气体交换膜组成,该膜具有与血液接触的超薄致密层,以防止在长期使用过程中血浆泄漏(铂立方NCVC)。离心泵(RotaFlow)无密封件。该系统的整个血液接触表面涂有一种新开发的肝素材料(东洋纺-NCVC涂层)。我们在山羊身上进行了静脉-动脉旁路手术,在不使用全身抗凝剂的情况下,ECMO系统运行了34天。氧合器未发生血浆泄漏,并保持了足够的气体交换性能。除了氧合器的外壳边缘外,在ECMO系统中几乎未观察到血栓形成。这种ECMO系统显示出在极少使用或不使用全身抗凝剂的情况下进行长期心肺支持的潜力。