在降低吸入氧化亚氮浓度后,双腔气管导管的支气管套囊迅速放气。

Rapid deflation of the bronchial cuff of the double-lumen tube after decreasing the concentration of inspired nitrous oxide.

作者信息

Karasawa Fujio, Takita Akira, Takamatsu Isao, Mori Tomohisa, Oshima Takashi, Kawatani Yasushi

机构信息

Department of Anesthesiology, National Defense Medical College, Tokorozawa, Saitama 359-8513, Japan.

出版信息

Anesth Analg. 2002 Jul;95(1):238-42, table of contents. doi: 10.1097/00000539-200207000-00044.

Abstract

UNLABELLED

Deflationary phenomena of the endotracheal tube cuff may occur after inspired nitrous oxide (N(2)O) concentrations are reduced, but deflationary phenomena of the double-lumen tube (DLT) cuff have not been investigated. In this study, tracheal and bronchial cuffs of left-sided Mallinckrodt (Athlone, Ireland) DLTs were inflated with air, 40% N(2)O, or 67% N(2)O (Air, N40, or N67 groups, respectively) in 24 patients undergoing thoracic surgery; 40 min later, O(2) was substituted for N(2)O in some of the patients in the N40 group (N40-c group). Intracuff gas volumes, N(2)O concentrations, and cuff compliance were also measured. Both tracheal and bronchial cuff pressures significantly increased in the Air group but decreased in the N67 group. Neither pressure significantly changed in the N40 group, but both decreased in the N40-c group after terminating N(2)O anesthesia; the time required for bronchial cuff pressures to decrease by half (12.0 +/- 5.5 min) was less than that for tracheal cuff pressures (31.2 +/- 11.0 min, P < 0.01). The volume change in the N40-c group was not significantly different between the tracheal and bronchial cuffs, but tracheal cuff compliance was significantly higher than bronchial compliance. Therefore, filling DLT cuffs with 40% N(2)O stabilizes cuff pressure during anesthesia with 67% N(2)O, but bronchial cuffs deflate more quickly than tracheal cuffs after cessation of N(2)O administration through smaller compliance.

IMPLICATIONS

We demonstrated that after cessation of nitrous oxide (N(2)O) administration, bronchial N(2)O-filled cuffs of the double-lumen tube deflate more rapidly than tracheal cuffs. To avoid insufficient separation of the lungs by the bronchial cuff, a frequent check of the cuff pressure is recommended after the inspired N(2)O concentration is decreased.

摘要

未标注

吸入氧化亚氮(N₂O)浓度降低后,气管导管套囊可能会出现放气现象,但双腔气管导管(DLT)套囊的放气现象尚未得到研究。在本研究中,对24例接受胸外科手术的患者,将左侧Mallinckrodt(爱尔兰阿斯隆)DLT的气管和支气管套囊分别用空气、40% N₂O或67% N₂O充气(分别为空气组、N40组或N67组);40分钟后,N40组的部分患者用O₂替代N₂O(N40-c组)。还测量了套囊内气体体积、N₂O浓度和套囊顺应性。空气组的气管和支气管套囊压力均显著升高,而N67组则降低。N40组的压力均无显著变化,但在终止N₂O麻醉后,N40-c组的压力均降低;支气管套囊压力降至一半所需的时间(12.0±5.5分钟)短于气管套囊压力所需时间(31.2±11.0分钟,P<0.01)。N40-c组气管和支气管套囊的体积变化无显著差异,但气管套囊顺应性显著高于支气管顺应性。因此,用40% N₂O充盈DLT套囊可在67% N₂O麻醉期间稳定套囊压力,但停止给予N₂O后,支气管套囊因顺应性较小而比气管套囊放气更快。

启示

我们证明,停止给予氧化亚氮(N₂O)后,双腔气管导管中充有N₂O的支气管套囊比气管套囊放气更快。为避免支气管套囊导致肺隔离不充分,建议在吸入N₂O浓度降低后频繁检查套囊压力。