Nogueira Iris P, Biggs Peter J
Department of Radiation Oncology, Massachusetts General Hospital, Harvard Medical School, Boston 02114, USA.
Health Phys. 2002 Aug;83(2):255-60. doi: 10.1097/00004032-200208000-00010.
Measurements have been made of the transmission factors through lead for scattered radiation produced by 4, 6, and 10 MV bremsstrahlung x-ray beams incident on a polystyrene phantom; these measurements cover the range of scattering angles between 30 degrees and 135 degrees. The results show that the tenth value layer (TVL) for scattered radiation decreases sharply with increasing scattering angle and increases slightly with increasing beam energy, although at large scattering angles there is little energy dependence. TVLs range from 3.5 cm to 0.3 cm for 4 MV at scattering angles between 30 degrees and 135 degrees, from 3.8 cm to 0.6 cm for 6 MV, and from 4.2 cm to 0.7 cm for 10 MV, respectively, at scattering angles between 30 degrees and 120 degrees. Monte Carlo calculations, performed at 4 MV to simulate the transmission of scattered radiation, are in good agreement with the experimental measurements.
已对4、6和10兆伏轫致辐射X射线束入射到聚苯乙烯体模上产生的散射辐射通过铅的传输因子进行了测量;这些测量涵盖了30度至135度的散射角范围。结果表明,散射辐射的第十值层(TVL)随散射角的增加而急剧减小,并随束流能量的增加而略有增加,不过在大散射角时能量依赖性很小。在30度至135度的散射角下,4兆伏时TVL范围为3.5厘米至0.3厘米,6兆伏时为3.8厘米至0.6厘米,10兆伏时为4.2厘米至0.7厘米,分别对应30度至120度的散射角。在4兆伏下进行的用于模拟散射辐射传输的蒙特卡罗计算与实验测量结果吻合良好。