Matching approved "nondedicated" hardware to obtain biventricular pacing and defibrillation: feasibility and troubleshooting.

作者信息

Kanagaratnam Logan, Pavia Stephen, Schweikert Robert, Marrouche Nassir, Lam Cathy, Abreu Miguel, Ching Elizabeth, Chung Mina, Saliba Walid, Niebauer Mark, Wilkoff Bruce, Tchou Patrick, Natale Andrea

机构信息

Department of Cardiology, Cleveland Clinic Foundation, Ohio 44195, USA.

出版信息

Pacing Clin Electrophysiol. 2002 Jul;25(7):1066-71. doi: 10.1046/j.1460-9592.2002.01066.x.

Abstract

Biventricular ICDs may offer increased benefit for patients with severe congestive heart failure and ventricular arrhythmia. Currently there are no approved dedicated biventricular ICDs available. Twenty-one consecutive patients who had approved nondedicated hardware implanted for biventricular pacing and defibrillation were included in this study. All device therapies were evaluated using stored electrograms. During mean follow-up at 13 +/- 7 months, 8 (36%) patients had inappropriate shocks. Ventricular fibrillation therapy was delivered for slow ventricular tachycardia because of double counting in two patients. In one patient, AV nodal reentrant tachycardia below detection rate cut off triggered device therapy because of ventricular double counting. Sinus tachycardia or premature atrial contraction initiating AV conduction and ventricular double counting resulted in shocks in five patients. The number of shocks per patient ranged from 1 to 64. Two patients required transient disconnection of the LV lead and subsequent ICD generator replacement for premature battery depletion. Two patients required AV junction ablation and three needed slow pathway ablation. Two patients were treated by upgrading to a device that was capable of a higher atrial tracking rate. The patients with impaired AV conduction or constant ventricular pacing did not have inappropriate therapy for sinus tachycardia or supraventricular arrhythmia. Use of conventional nondedicated hardware for biventricular pacer/defibrillator is feasible but should be considered only in patients with poor AV node function or less likely to require antitachycardic therapy, to avoid ICD double counting of ventricular sensed events and consequent high incidence of inappropriate therapies.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验