Metzler S D, Bowsher J E, Greer K L, Jaszczak R J
Duke University Medical Center, P. O. Box 3949, Divison of Nuclear Medicine, Durham, NC 27710, USA.
IEEE Trans Med Imaging. 2002 Aug;21(8):878-87. doi: 10.1109/tmi.2002.803129.
Pinhole collimators are widely used to image small organs and animals. The pinhole response function (PRF) of knife-edge pinhole collimators has been estimated previously using geometric constructions without considering penetration and using "roll-off" models that employ an exponential model for the flux. An analytic expression for the PRF on the imaging plane that includes the effect of aperture penetration is derived in this paper by calculating the flux for photons passing through the aperture and those passing through the attenuating material. The PRF is then used to approximate the angular-dependent root-mean-square resolution in the directions parallel and perpendicular to the tilt of the point source. The corresponding aspect ratio is then obtained. The formulas are then compared with experimental data.
针孔准直器被广泛用于对小型器官和动物进行成像。此前,已经使用几何结构估计了刀刃式针孔准直器的针孔响应函数(PRF),但未考虑穿透效应,并且使用了对通量采用指数模型的“滚降”模型。本文通过计算穿过孔径的光子和穿过衰减材料的光子的通量,推导出了成像平面上包含孔径穿透效应的PRF解析表达式。然后,利用PRF来近似点源倾斜方向平行和垂直方向上与角度相关的均方根分辨率。接着得到相应的纵横比。最后将这些公式与实验数据进行比较。