Yamagata Yukari, Shimizu Kayoko, Nakamura Kenji, Henmi Fumiyo, Satomura Shinji, Matsuura Shuji, Tanaka Masatoshi
Osaka Research Laboratories, Wako Pure Chemical Industries, Ltd, 6-1 Takada-cho, Amagasaki, Hyogo 661-0963, Japan.
Clin Chim Acta. 2003 Jan;327(1-2):59-67. doi: 10.1016/s0009-8981(02)00305-4.
Lens culinaris agglutinin (LCA)-reactive alpha-fetoprotein (AFP-L3) percentage of total AFP concentration [(AFP-L3/total AFP)x100] has been used as an effective marker for earlier diagnosis, for assessment of therapeutic effects and for predicting the prognosis of hepatocellular carcinoma (HCC).
A new clinical automatic analyzer, the "LiBASys", and an assay kit for the simultaneous determination of AFP-L3 percentage and concentration of AFP in human serum were developed. LiBASys performed automatic re-measurement and also met stat samples. Both AFP-L3 percentage and AFP concentration were calculated simultaneously and printed out continuously every 3.4 min per test.
A linear dose-response relationship was observed up to 1000 ng/ml AFP concentration. The detection limit and functional sensitivity of LCA-nonreactive AFP (AFP-L1) and AFP-L3 concentrations were 0.4 and 0.8 ng/ml, and 0.4 and 1.0 ng/ml, respectively. Interassay CVs of AFP-L3 percentage and AFP concentration ranged from 2.8% to 13.4% and 2.6% to 4.6%, respectively. Assay results exhibited good correlations with those of commercially available lectin-affinity electrophoresis (r=0.984) for AFP-L3 percentage and with the RIA method (r=0.999) for AFP concentration.
This method simultaneously yields both qualitative and quantitative results for AFP-L3 percentage and AFP concentration.
Lens culinaris凝集素(LCA)反应性甲胎蛋白(AFP-L3)占总AFP浓度的百分比[(AFP-L3/总AFP)×100]已被用作肝细胞癌(HCC)早期诊断、治疗效果评估及预后预测的有效标志物。
研发了一种新型临床自动分析仪“LiBASys”以及用于同时测定人血清中AFP-L3百分比和AFP浓度的检测试剂盒。LiBASys可进行自动复测,也能满足急诊样本检测。每次检测时,AFP-L3百分比和AFP浓度会同时计算,并每3.4分钟连续打印一次结果。
在AFP浓度达1000 ng/ml时观察到线性剂量反应关系。LCA非反应性AFP(AFP-L1)和AFP-L3浓度的检测限及功能灵敏度分别为0.4和0.8 ng/ml,以及0.4和1.0 ng/ml。AFP-L3百分比和AFP浓度的批间变异系数分别为2.8%至13.4%和2.6%至4.6%。检测结果与市售凝集素亲和电泳法(AFP-L3百分比,r = 0.984)及RIA法(AFP浓度,r = 0.999)的结果具有良好相关性。
该方法可同时得出AFP-L3百分比和AFP浓度的定性和定量结果。