Lowe David, Hoare Tim
Department of Otolaryngology, Bedford Hospital NHS Trust, Bedford, UK.
J Laryngol Otol. 2002 Nov;116(11):914-6. doi: 10.1258/00222150260369444.
The Doughty split tongue blade is used routinely with the Boyle Davis mouth gag during tonsillectomy. If the width of the slot in the blade is excessive, then the endotracheal tube may herniate through the slot in the blade and become fixed and inseparable from the blade. We report two cases of paediatric tonsillectomy where this potentially serious complication occurred. The slot widths of disposable and reusable Doughty blades were measured and specifications were obtained from several manufacturers. All disposable paediatric Doughty blades and some of the reusable blades were found to have excessively and potentially dangerously wide slots. Compression testing of endotracheal tubes by the manufacturer demonstrated marked increases in deformability on warming to body temperature. Ensuring that the slot width of the Doughty blade is not excessive for the endotracheal tube used is essential and we give recommendations.
在扁桃体切除术中,多蒂(Doughty)开口器通常与博伊尔·戴维斯(Boyle Davis)开口器一起使用。如果开口器中的狭槽宽度过大,气管内导管可能会从开口器的狭槽中疝出,并与开口器固定在一起且无法分离。我们报告了两例小儿扁桃体切除术中发生这种潜在严重并发症的病例。测量了一次性和可重复使用的多蒂开口器的狭槽宽度,并从多家制造商处获取了规格。发现所有一次性小儿多蒂开口器和一些可重复使用的开口器的狭槽宽度都过大,存在潜在危险。制造商对气管内导管进行的压缩测试表明,在升温至体温时,导管的可变形性显著增加。确保多蒂开口器的狭槽宽度对于所使用的气管内导管不过大至关重要,我们给出了相关建议。