A new 90Sr/90Y catheter-based system, the Beta-Cath 3.5F system, was recently introduced into the clinic for intravascular brachytherapy. This new system has a smaller diameter of delivery catheter (3.5 French) compared to the original Beta-Cath system (5 French catheter). The new source/seed (3.5F source) has a different design from the original (5F) source. In this paper, we report the EGSnrc Monte Carlo results of two-dimensional dose rate distributions in water as well as the TG-60 dose parameters for the new 3.5F source and the original 5F source. A complete set of the dosimetric data for each individual seed is tabulated. A detailed dosimetric comparison was carried out for the two sources. The comparison shows that differences between the two sources are less than 5% in the region of interest for intravascular brachytherapy. Higher differences (up to 10%) are seen in the near-source region, which is mainly inside the delivery catheter. Based on the present calculation and comparison, we conclude that the new Beta-Cath 3.5F system is dosimetrically equivalent to the original Beta-Cath (5F) system. Therefore, clinical dosimetry and the practice of using both the systems can be treated similarly.
一种新型的基于90Sr/90Y导管的系统——Beta-Cath 3.5F系统,最近已引入临床用于血管内近距离放射治疗。与原始的Beta-Cath系统(5F导管)相比,这种新系统的输送导管直径更小(3.5法国规格)。新的源/籽源(3.5F源)与原始的(5F)源设计不同。在本文中,我们报告了新的3.5F源和原始5F源在水中的二维剂量率分布的EGSnrc蒙特卡罗结果以及TG-60剂量参数。列出了每个单独籽源的完整剂量学数据。对这两种源进行了详细的剂量学比较。比较表明,在血管内近距离放射治疗的感兴趣区域,两种源之间的差异小于5%。在靠近源的区域,即主要在输送导管内部,差异更大(高达10%)。基于目前的计算和比较,我们得出结论,新的Beta-Cath 3.5F系统在剂量学上等同于原始的Beta-Cath(5F)系统。因此,临床剂量学以及使用这两种系统的实践可以类似地进行处理。