Malplacement of endocardial pacemaker electrodes in the middle cardiac vein.

作者信息

Kemp A, Kjersgaard Johansen J, Kjaergaard E

出版信息

Acta Med Scand. 1976;199(1-2):7-11. doi: 10.1111/j.0954-6820.1976.tb06684.x.

Abstract

Endocardial transvenous pacemaker electrodes have been implanted in 246 patients owing to symptomatic bradycardia. The electrode was malplaced in the middle cardiac vein in 12 patients. All the patients were on stable ventricular pacing at the time of implantation. The electrodes were still in function in six patients after 5-96 months (average 55). The electrode was replaced in three patients owing to the occurrence of exit block. Three patients died during the period under study. Right-sided bundle branch block (RBBB) configuration of paced QRS complexes was observed on the ECG when the electrode was malplaced in the middle cardiac vein. As RBBB only occurs with malplacement, the ECG can be employed to ensure correct positioning of the electrode.

摘要

相似文献

1
Malplacement of endocardial pacemaker electrodes in the middle cardiac vein.
Acta Med Scand. 1976;199(1-2):7-11. doi: 10.1111/j.0954-6820.1976.tb06684.x.
3
Transvenous pacemakers in clinical practice.
Med Clin North Am. 1973 Jul;57(4):1001-14. doi: 10.1016/s0025-7125(16)32247-7.
5
Sutureless epicardial pacemaker lead: a satisfactory preliminary experience.
Chest. 1975 May;67(5):564-7. doi: 10.1378/chest.67.5.564.
6
Temporary transvenous pacemaker therapy: an analysis of complications.
Chest. 1973 Nov;64(5):604-8. doi: 10.1378/chest.64.5.604.
7
Inadvertent Left Ventricle Endocardial or Uncomplicated Right Ventricular Pacing: How to Differentiate in the Emergency Department.
J Emerg Med. 2018 May;54(5):e91-e95. doi: 10.1016/j.jemermed.2018.01.034. Epub 2018 Mar 6.
9
Cardiac pacing in children.
Cardiovasc Clin. 1983;14(2):209-26.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验