Influence of leg position and environmental temperature on segmental volume expansion during venous occlusion plethysmography.

作者信息

Jorfeldt Lennart, Vedung Torbjörn, Forsström Elisabeth, Henriksson Jan

机构信息

Department of Surgical Sciences, Division of Clinical Physiology, Karolinska Hospital, SE-171 76 Stockholm, Sweden.

出版信息

Clin Sci (Lond). 2003 Jun;104(6):599-605. doi: 10.1042/CS20020257.

Abstract

Blood flow determinations by venous occlusion plethysmography applying the strain-gauge technique are frequently used. A problem with the strain-gauge technique is that the relationship between venous volume and transmural pressure is not linear and, furthermore, changes with the sympathetic tone. The present study tests the hypothesis that these factors lead to a redistribution of venous blood, which may impair the accuracy of the technique. The relative volume expansion rates of four leg segments were studied with the leg in different positions and at disparate temperatures, thereby inducing varying venous pressures and sympathetic tone ( n =6). With elevated leg and relaxed veins (at 50 degrees C), the distal thigh showed a relatively low expansion rate (25.8+/-4.5 ml.min(-1).l(-1)), whereas values in the calf segments were higher (34.5-39.0 ml.min(-1).l(-1)). With lower initial transmural pressure, calf segments can increase their volume much more during occlusion compared with the distal thigh. In a higher transmural pressure region (lowered leg), the difference in compliance between limb segments is less. In this case, compliance and volume expansion rate was higher in the distal thigh (14.2, 13.5 and 22.2 ml.min(-1).l(-1) at 10, 20 and 50 degrees C respectively) than in the calf segments (for the distal calf: 6.4, 7.7 and 16.2 ml.min(-1).l(-1) respectively). There was a significant interaction ( P <0.001) between temperature and leg position, indicating a higher degree of sympathetic vasoactivity in the calf. It is concluded that blood flow determination by strain-gauge plethysmography is less accurate, due to a potential redistribution of the venous blood. Therefore possible influences of variations in sympathetic tone and venous pressure must be considered even in intra-individual comparisons, especially in interventional studies.

摘要

通过应用应变片技术的静脉阻塞体积描记法来测定血流量是常用的方法。应变片技术存在一个问题,即静脉容量与跨壁压力之间的关系并非线性,而且会随交感神经张力而变化。本研究检验了这样一个假设,即这些因素会导致静脉血重新分布,这可能会损害该技术的准确性。在腿部处于不同位置和不同温度下,研究了四个腿部节段的相对容量扩张率,从而诱导出不同的静脉压力和交感神经张力(n = 6)。当腿部抬高且静脉松弛时(50摄氏度),大腿远端的扩张率相对较低(25.8±4.5毫升·分钟⁻¹·升⁻¹),而小腿节段的值更高(34.5 - 39.0毫升·分钟⁻¹·升⁻¹)。与大腿远端相比,在初始跨壁压力较低时,小腿节段在阻塞期间能够使自身容量增加更多。在跨壁压力较高的区域(腿部下垂),肢体节段之间的顺应性差异较小。在这种情况下,大腿远端的顺应性和容量扩张率更高(在10、20和50摄氏度时分别为14.2、13.5和22.2毫升·分钟⁻¹·升⁻¹),高于小腿节段(小腿远端分别为6.4、7.7和16.2毫升·分钟⁻¹·升⁻¹)。温度和腿部位置之间存在显著交互作用(P < 0.001),表明小腿的交感神经血管活性程度更高。得出的结论是,由于静脉血可能会重新分布,通过应变片体积描记法测定血流量的准确性较低。因此,即使在个体内部比较中,尤其是在干预性研究中,也必须考虑交感神经张力和静脉压力变化的可能影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验