MacDonald C A, Gibson W M, Peppler W W
Center for X-ray Optics, Department of Physics, University at Albany, Albany, NY 12222, USA.
Technol Cancer Res Treat. 2002 Apr;1(2):111-7. doi: 10.1177/153303460200100203.
Polycapillary x-ray optics provide an innovative new way to control x-ray beams. Placing these optics after the object to be imaged provides very efficient rejection of Compton scatter, while allowing image magnification without loss of resolution, image demagnification, or image shaping to match with digital detectors. Measured scatter rejection optics had primary transmissions greater than 50% and scatter transmission of less than 1%. For a 5-cm thick Lucite phantom, this resulted in a contrast enhancement of nearly a factor of two at 20 keV and three at 40 keV. The magnification from the tapered capillary optics improved the MTF at all frequencies out to 1.8 times the original system resolution. Increases below the system resolution are most important because clinically relevant structures generally occupy lower spatial frequencies. Alternatively, placing a collimating optic and diffracting crystal before the patient provides sufficient monochromatic beam intensity for medical imaging. Contrast, resolution, and intensity measurements were performed with both high and low angular acceptance crystals. At 8 keV, contrast enhancement was a factor of 5 relative to the polychromatic case, in good agreement with theoretical values. At 17.5 keV, monochromatic subject contrast was more than a factor of 2 times greater than the conventional polychromatic contrast. An additional factor of two increase in contrast, for a total factor of four, is expected from the removal of scatter in a large beam clinical system. The measured angular resolution after the crystal was 0.4 mrad for a silicon crystal.
多毛细管X射线光学器件提供了一种控制X射线束的创新方法。将这些光学器件放置在待成像物体之后,可以非常有效地抑制康普顿散射,同时实现图像放大而不损失分辨率、图像缩小或图像整形,以匹配数字探测器。经测量,散射抑制光学器件的主要透射率大于50%,散射透射率小于1%。对于一个5厘米厚的有机玻璃模型,这在20千电子伏时使对比度增强了近两倍,在40千电子伏时增强了三倍。锥形毛细管光学器件的放大作用提高了所有频率下的调制传递函数(MTF),最高可达原始系统分辨率的1.8倍。低于系统分辨率的增加最为重要,因为临床相关结构通常占据较低的空间频率。或者,在患者之前放置一个准直光学器件和衍射晶体可为医学成像提供足够的单色光束强度。使用高角度接受晶体和低角度接受晶体进行了对比度、分辨率和强度测量。在8千电子伏时,相对于多色情况,对比度增强了5倍,与理论值吻合良好。在17.5千电子伏时,单色物体对比度比传统多色对比度大两倍多。预计在大型光束临床系统中通过去除散射,对比度还会再增加两倍,总共增加四倍。对于硅晶体,晶体后的测量角分辨率为0.4毫弧度。