Suppr超能文献

Simultaneous measurement of phenylalanine and tyrosine in phenylketonuric plasma and dried blood by high-performance liquid chromatography.

作者信息

Dale Y, Mackey V, Mushi R, Nyanda A, Maleque M, Ike J

机构信息

Department of Biology, Fisk University, 1000 17th Avenue North, Nashville, TN 37208, USA.

出版信息

J Chromatogr B Analyt Technol Biomed Life Sci. 2003 May 5;788(1):1-8. doi: 10.1016/s1570-0232(02)01005-x.

Abstract

Phenylketonuria (PKU) is a disorder characterized by an interruption in the conversion of phenylalanine to tyrosine, a reaction catalyzed by phenylalanine hydroxylase (PAH). Animal models of PKU used in this study were induced by daily subcutaneous injections of pups with alpha-methylphenylalanine plus phenylalanine in utero and postnatally from day 4 to day 14. Dry blood and plasma were utilized to measure phenylalanine concentration in PKU rats. The results indicated that the concentration of phenylalanine is higher and more stable in plasma than dry blood. Precolumn derivatization of dried blood and plasma free amino acids were conducted with phenylisothiocyanate (PITC). The phenylthiocarbamyl (PTC) derivatives were separated on a reversed-phase C-18 column (15 cm x 4.6 mm). A gradient high-performance liquid chromatography method with two eluents, 0.1 M sodium acetate buffer and 100% acetonitrile was developed to facilitate the separation of nine amino acids within 11 min. Tyrosine and phenylalanine eluted the column at 5.4 and 9.4 min, respectively. This method provides a quick and reliable technique for neonatal screening.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验