Suppr超能文献

用于混合型正电子发射断层显像(PET)成像的平行板准直器设计综述。

Survey of parallel slat collimator designs for hybrid PET imaging.

作者信息

Rust T C, Kadrmas D J

机构信息

Medical Imaging Research Laboratory, Department of Radiology, University of Utah, CAMT, 729 Arapeen Drive, Salt Lake City, UT 84108-1218, USA.

出版信息

Phys Med Biol. 2003 Mar 21;48(6):N97-104. doi: 10.1088/0031-9155/48/6/402.

Abstract

Hybrid PET gamma cameras with coincidence detection electronics are commonly equipped with parallel slat collimators in order to reduce detection of singles and scattered photons, and create a pseudo-2D imaging geometry. The objective of this work was to survey a broad range of parallel slat collimator designs using a series of Monte Carlo simulated PET acquisitions. Collimator properties including septal height, septal thickness and pitch were independently examined over a wide range of values. Simulations were performed for hybrid PET imaging of a long cylindrical phantom uniformly filled with water and radioactivity. The performance for each collimator design was evaluated in terms of the trues-to-singles ratio, scatter fraction, and noise equivalent count rate for a wide range of camera trigger rates. Results indicate that increasing septal height offers the biggest performance gain. Septal thickness should be at least 0.5 mm, and should be optimized in conjunction with pitch to obtain the best performance. This survey provides the groundwork necessary for optimizing slat collimators, and provides a starting point for investigating new slat collimator designs.

摘要

带有符合探测电子设备的混合型正电子发射断层显像(PET)γ相机通常配备平行板准直器,以减少单光子和散射光子的探测,并创建一种伪二维成像几何结构。这项工作的目的是通过一系列蒙特卡罗模拟的PET采集来研究广泛的平行板准直器设计。在很宽的值范围内独立检查准直器特性,包括隔板高度、隔板厚度和间距。对均匀填充水和放射性物质的长圆柱形体模进行混合型PET成像模拟。针对各种相机触发率,根据真符合计数与单光子计数比、散射分数和噪声等效计数率评估每种准直器设计的性能。结果表明,增加隔板高度可带来最大的性能提升。隔板厚度应至少为0.5毫米,并应与间距一起优化以获得最佳性能。这项研究为准直器的优化提供了必要的基础,并为研究新的准直器设计提供了一个起点。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验